MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/opth.s476868

Медицинская библиотека · Clinical ophthalmology (Auckland, N.Z.) · 2024

Evaluation of Sensitivity and Specificity of Acuity 360 Telemedicine Vision Screening System.

Binder Nicholas R, Bloom Jeffrey M, Balushian Haftevani Mineh, Duncan Stephen, Lowry Nathan Christopher, Chang Tom S

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Our aim is to evaluate the sensitivity and specificity of Acuity 360 telemedicine system, as compared to in-person clinic examination, in identifying clinically significant eye disease. Acuity 360 is a combination of commercially available ocular imaging devices used together to provide a comprehensive evaluation of the structures and diseases of the eye.

METHODS

Observational cross-sectional study of consecutively examined patients where 19 remote examiners analyzed 80 patients using Acuity 360 images. The examiners' diagnoses were compared to the diagnosis obtained from in-person clinic examination of all the patients. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), false-negative rate, and inter-rater agreement were calculated.

RESULTS

Compared to in-person clinic examination, the sensitivity, specificity, PPV, NPV, and false-negative rate for the optical coherence tomography (OCT) of optic nerve were 95.8%, 98.5%, 86.8%, 99.6%, 0.39% and for the combined retina examination (OCT of the macula and widefield fundus photography) were 93.4%, 88.5%, 80.2%, 96.4%, 2.19%, respectively. The median inter-rater agreement the OCT of optic nerve and the combined retina examination were each 95%.

CONCLUSION

The Acuity 360 telemedicine system has a low false-negative rate and is highly sensitive and specific when compared to an in-person clinic examination. It can determine the necessity for specialist referral and triage the patients that require urgent treatment. High inter-rater agreement shows that it is effective with minimal variability in analyzing the Acuity 360 images by remote examiners.

Для цитирования: Binder Nicholas R, Bloom Jeffrey M, Balushian Haftevani Mineh, Duncan Stephen, Lowry Nathan Christopher, Chang Tom S. Evaluation of Sensitivity and Specificity of Acuity 360 Telemedicine Vision Screening System.. Clinical ophthalmology (Auckland, N.Z.). 2024. DOI: 10.2147/opth.s476868.