An in silico guide for ventriculo-ventricular delay programming for left bundle branch-optimized cardiac resynchronization therapy.

Strocchi Marina, Samways Jack W, Naraen Akriti, Ali Nadine, Shun-Shin Matthew J, Gillette Karli, Rinaldi Christopher Aldo, Arnold Ahran D, Plank Gernot, Vigmond Edward J, Whinnett Zachary I, Niederer Steven A

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Left bundle branch pacing (LBBP)-optimized cardiac resynchronization therapy (LOT-CRT) can improve left ventricular (LV) activation when LBBP alone or conventional biventricular pacing are ineffective. However, the optimal programming settings for ventriculo-ventricular delay (VVD) for LOT-CRT are unknown. We aim to investigate how to optimally program VVD for LOT-CRT in the presence of various LV conduction substrates using computational modelling.

METHODS AND

RESULTS

We simulated ventricular activation on 24 anatomies and validated the model against clinical data. Diffuse LV conduction system and intra-myocardial delay were simulated by slowing the conduction velocity of the LV His-Purkinje system and myocardium, respectively, alone or in combination with proximal left bundle branch block (LBBB). We simulated LOT-CRT with selective or myocardial capture (LV septal pacing, LVSP) with VVD ranging between -100 ms (LBBP/LVSP ahead) and +100 ms [LV epicardial lead (LVepiP), ahead]. Response was quantified with 95% LV activation times (LVAT95). In the presence of diffuse LV conduction system delay, the optimal VVD for LOT-CRT was always negative (LBBB: -42.5 ± 6.6 ms; no LBBB: -36.2 ± 5.6 ms), as delivering LBBP ahead of LVepiP compensates for the slow LV His-Purkinje. In the presence of LV intra-myocardial disease, the shortest LVAT95 with LOT-CRT was achieved by pacing the coronary sinus LV first (optimal VVD for LBBB: 23.3 ± 8.5 ms; no LBBB: 79.2 ± 18.0 ms). The type of capture for LOT-CRT affected the optimal VVD, with myocardial capture favouring negative VVDs (LVSP ahead).

CONCLUSION

The optimal VVD for LOT-CRT depends on the mechanism of delayed LV activation and type of capture achieved, highlighting the importance of VVD optimization.

Для цитирования: Strocchi Marina, Samways Jack W, Naraen Akriti, Ali Nadine, Shun-Shin Matthew J, Gillette Karli, Rinaldi Christopher Aldo, Arnold Ahran D, Plank Gernot, Vigmond Edward J, Whinnett Zachary I, Niederer Steven A. An in silico guide for ventriculo-ventricular delay programming for left bundle branch-optimized cardiac resynchronization therapy.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2025. DOI: 10.1093/europace/euaf089.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.