Аннотация

PURPOSE: We developed a transurethral resection system comprising a uniquely designed resectoscope and high frequency wave generator. The obturator nerve is free of electrical stimulation during transurethral resection since high frequency current is delivered via a resection loop to the sheath of the resectoscope in saline irrigation. Preclinical verification of the system was performed in an animal model to ensure its efficacy and operational safety.

MATERIALS AND

METHODS

The swine bladder wall was transurethrally resected using the system in a saline environment. Results were subsequently compared with data on identical resection performed with the conventional system in sorbitol solution irrigation. Electrolyte contents were measured after resection for comparative evaluation with corresponding pre-resection data. Also, the depth of heat degeneration was measured in resected tissue.

RESULTS

No additional skills were required for bladder resection performed with the new system versus the conventional system. No lower limb adductor contraction was noted except minimal creeping during resection of a site close to the urethra. There were no apparent anomalies of blood electrolytes after transurethral resection. No difference was observed in the mean depth of heat degenerated tissues compared with the conventional system.

CONCLUSIONS

The newly developed transurethral resection system was effective in a saline irrigated environment, which inhibited the obturator nerve reflex when applied in an animal model. This finding suggests promising potential for the system as a safe and cost-effective alternative compared with conventional transurethral resection.

Для цитирования: Shiozawa Hiroaki, Aizawa Taku, Ito Takaaki, Miki Makoto. A new transurethral resection system: operating in saline environment precludes obturator nerve reflexes.. The Journal of urology. 2002. DOI: 10.1016/s0022-5347(05)64240-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.