The neochord mitral valve repair procedure: Numerical simulation of different neochords tensioning protocols.

Di Micco Luigi, Peruzzo Paolo, Colli Andrea, Burriesci Gaetano, Boso Daniela, Besola Laura, Gerosa Gino, Susin Francesca M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Transapical off-pump mitral valve repair with neochord implantation is an established technique for minimally-invasive intervention on mitral valve prolapse/flail. The procedure involves the positioning of artificial chords, whose length/tension is adjusted intraoperatively, adopting different methods based on the experience of the surgeon. This unsystematic approach occasionally leads to complications such as leaflet rupture and excessive/insufficient load on the neochords. In this study, finite element models of a generalized prolapsing mitral valve are used to verify the effect of two alternative tensioning approaches (AT - All together and 1-by-1 - one by one sequences) on the coaptation area and valve biomechanics, comparing results with a corresponding healthy configuration. The total force of about 1 N is exerted by the chords in both strategies, but the maximum stress and coaptation area are closer to those of the healthy configuration in the 1-by-1 sequence. However, the analysis also provides an explanation for the chords unloading in the 1-by-1 strategy observed in the clinical practice, and suggests an optimum tensioning methodology for NeoChord procedures. The study also reveals the potential power of the implemented numerical approach to serve as a tool for procedural planning, supporting the identification of the most suitable ventricular access site and the most effective stitching points for the artificial chords.

Для цитирования: Di Micco Luigi, Peruzzo Paolo, Colli Andrea, Burriesci Gaetano, Boso Daniela, Besola Laura, Gerosa Gino, Susin Francesca M. The neochord mitral valve repair procedure: Numerical simulation of different neochords tensioning protocols.. Medical engineering & physics. 2019. DOI: 10.1016/j.medengphy.2019.09.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.