Evaluation of the performance of the Abbott ARCHITECT HIV Ag/Ab Combo Assay.

Chavez Pollyanna, Wesolowski Laura, Patel Pragna, Delaney Kevin, Owen S Michele

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Worldwide, many countries test for HIV infection using combination assays that simultaneously detect p24 antigen and HIV antibodies. One such assay, the ARCHITECT(®) HIV Ag/Ab Combo Assay (ARCHITECT), has recently been approved by the Food and Drug Administration (FDA) for use in the United States.

OBJECTIVE

To evaluate the performance of ARCHITECT on well-characterized specimens from four CDC-funded studies.

STUDY

DESIGN

We evaluated 3386 HIV-infected, 7551 HIV-uninfected, and 58 acute HIV infection (AHI) specimens. HIV-infected specimens were repeatedly reactive by enzyme immunoassay (EIA) and Western blot (WB) or positive by nucleic acid amplification testing (NAAT). HIV-uninfected specimens were EIA- and NAAT-negative. AHI specimens were seronegative or indeterminate (using antibody-based EIAs, rapid tests or WB) and NAAT-positive. All specimens were de-identified and sent to Abbott Diagnostics for testing with ARCHITECT. ARCHITECT test results were compared to original study characterizations and were used to assess overall sensitivity and specificity and also sensitivity for AHI. ARCHITECT false-positive specimens with sufficient quantity were retested.

RESULTS

Based on results from the initial ARCHITECT test, sensitivity was 99.94% (95% confidence interval [CI]: 99.79, 99.99) and specificity was 98.78% (95% CI: 98.51-99.01). Repeat testing resulted in corrected specificity of 99.50% (95% CI: 99.31, 99.64). Also, 48 AHI specimens (83%) were detected by this screening assay.

CONCLUSION

The sensitivity and specificity of the ARCHITECT combination assay are very high and most AHIs were detected by the assay. Use of Ag/Ab combination assays may improve the number of AHIs identified relative to existing FDA-approved HIV-antibody only based serologic assays, particularly in high incidence populations.

Для цитирования: Chavez Pollyanna, Wesolowski Laura, Patel Pragna, Delaney Kevin, Owen S Michele. Evaluation of the performance of the Abbott ARCHITECT HIV Ag/Ab Combo Assay.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2011. DOI: 10.1016/j.jcv.2011.09.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.