Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.

Abicca Irene, Schiano-Lomoriello Domenico, Gilardi Marta, Giannini Daniela, Dinu Valentin, Coutinho Catarina, Savini Giacomo

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus.

METHODS

Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices.

RESULTS

Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry.

CONCLUSIONS

The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. [J Refract Surg. 2024;40(7):e445-e452.].

Для цитирования: Abicca Irene, Schiano-Lomoriello Domenico, Gilardi Marta, Giannini Daniela, Dinu Valentin, Coutinho Catarina, Savini Giacomo. Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2024. DOI: 10.3928/1081597x-20240514-03.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.