MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/hae.70007

Медицинская библиотека · Haemophilia : the official journal of the World Federation of Hemophilia · 2025

Factor VIII Activity and Factor VIII Inhibitors Can Be Measured Accurately in Plasma Containing Mim8 by Using Specific Chromogenic Assays.

Pickering William, Robinson Mary, Cogswell Caroline, Clausen Wan Hui Ong, Weldingh Karin Nana, Ezban Mirella

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Patients with Haemophilia A treated with Mim8 may require concomitant Factor VIII (FVIII) replacement; hence, accurate assessment of FVIII activity (FVIII:C) and FVIII inhibitors in the presence of Mim8 is important.

AIM: Assess which chromogenic substrate assays (CSAs) accurately monitor FVIII:C and FVIII inhibitor levels in the presence of Mim8.

METHODS

FVIII and Mim8 were spiked into congenital FVIII-deficient plasma to obtain mixtures containing various FVIII:C levels (0-100 IU/dL) and Mim8 concentrations (0, 3, 6 and 12 µg/mL). Five CSAs using different activated Factor IX and X sources (bovine, bovine-human or human) were used to measure FVIII:C. Bovine and bovine-human CSAs were then used to measure FVIII:C of standard half-life (SHL) and extended half-life (EHL) FVIII products. FVIII inhibitor levels were assessed using two different bovine CSAs in congenital FVIII-deficient plasma spiked with various Mim8 concentrations (5, 10, 20 and 40 µg/mL) and FVIII inhibitor levels (0.2, 1.0 and 4.8 BU).

RESULTS

High levels of interference were observed using human CSAs. Bovine CSAs accurately measured FVIII:C of SHL and EHL FVIII products in the presence of Mim8 without interference. In bovine-human CSAs, interference was observed at 5 IU/dL FVIII, increasing up to four-fold with increasing Mim8 levels. FVIII inhibitor levels were accurately measured using bovine CSAs without Mim8 interference.

CONCLUSION

FVIII:C of SHL and EHL products and FVIII inhibitor levels can be accurately monitored in the presence of Mim8 using bovine CSAs at all FVIII levels, and bovine-human CSAs at FVIII concentrations >20 IU/dL.

Для цитирования: Pickering William, Robinson Mary, Cogswell Caroline, Clausen Wan Hui Ong, Weldingh Karin Nana, Ezban Mirella. Factor VIII Activity and Factor VIII Inhibitors Can Be Measured Accurately in Plasma Containing Mim8 by Using Specific Chromogenic Assays.. Haemophilia : the official journal of the World Federation of Hemophilia. 2025. DOI: 10.1111/hae.70007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.