MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4244/eij-d-22-00289

Медицинская библиотека · EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2022

Haemodynamic performance and clinical outcomes of transcatheter aortic valve replacement with the self-expanding ACURATE neo2.

Scotti Andrea, Pagnesi Matteo, Kim Won-Keun, Schäfer Ulrich, Barbanti Marco, Costa Giuliano, Baggio Sara, Casenghi Matteo, De Marco Federico, Vanhaverbeke Maarten, Sondergaard Lars, Wolf Alexander, Schofer Joachim, Ancona Marco Bruno, Montorfano Matteo, Kornowski Ran, Assa Hana Vaknin, Toggweiler Stefan, Ielasi Alfonso, Hildick-Smith David, Windecker Stephan, Schmidt Albrecht, Buono Andrea, Maffeo Diego, Siqueira Dimytri, Giannini Francesco, Adamo Marianna, Massussi Mauro, Wood David A, Sinning Jan-Malte, Van Der Heyden Jan, van Ginkel Dirk-Jan, Van Mieghem Nicholas, Veulemans Verena, Mylotte Darren, Tzalamouras Vasileios, Taramasso Maurizio, Estévez-Loureiro Rodrigo, Colombo Antonio, Mangieri Antonio, Latib Azeem

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Transcatheter aortic valve replacement (TAVR) with the ACURATE neo device has been associated with a non-negligible incidence of paravalvular aortic regurgitation (AR). The new-generation ACURATE neo2 has been designed to mitigate this limitation.

AIMS: The aim of the study was to compare TAVR with the ACURATE neo and neo2 devices.

METHODS

The NEOPRO and NEOPRO-2 registries retrospectively included patients undergoing transfemoral TAVR with self-expanding valves at 24 and 20 centres, respectively. Patients receiving the ACURATE neo and neo2 devices (from January 2012 to December 2021) were included in this study. Predischarge and 30-day VARC-3 defined outcomes were evaluated. The primary endpoint was predischarge moderate or severe paravalvular AR. Subgroup analyses per degree of aortic valve calcification were performed.

RESULTS

A total of 2,026 patients (neo: 1,263, neo2: 763) were included. Predischarge moderate or severe paravalvular AR was less frequent for the neo2 group (2% vs 5%; p<0.001), resulting in higher VARC-3 intended valve performance (96% vs 90%; p<0.001). Furthermore, more patients receiving the neo2 had none/trace paravalvular AR (59% vs 38%; p<0.001). The reduction in paravalvular AR with neo2 was mainly observed with heavy aortic valve calcification. New pacemaker implantation and VARC-3 technical and device success rates were similar between the 2 groups; there were more frequent vascular and bleeding complications for the neo device. Similar 1-year survival was detected after TAVR (neo2: 90% vs neo: 87%; p=0.14).

CONCLUSIONS

TAVR with the ACURATE neo2 device was associated with a lower prevalence of moderate or severe paravalvular AR and more patients with none/trace paravalvular AR. This difference was particularly evident with heavy aortic valve calcification.

Для цитирования: Scotti Andrea, Pagnesi Matteo, Kim Won-Keun, Schäfer Ulrich, Barbanti Marco, Costa Giuliano, Baggio Sara, Casenghi Matteo, De Marco Federico, Vanhaverbeke Maarten, Sondergaard Lars, Wolf Alexander, Schofer Joachim, Ancona Marco Bruno, Montorfano Matteo, Kornowski Ran, Assa Hana Vaknin, Toggweiler Stefan, Ielasi Alfonso, Hildick-Smith David, Windecker Stephan, Schmidt Albrecht, Buono Andrea, Maffeo Diego, Siqueira Dimytri, Giannini Francesco, Adamo Marianna, Massussi Mauro, Wood David A, Sinning Jan-Malte, Van Der Heyden Jan, van Ginkel Dirk-Jan, Van Mieghem Nicholas, Veulemans Verena, Mylotte Darren, Tzalamouras Vasileios, Taramasso Maurizio, Estévez-Loureiro Rodrigo, Colombo Antonio, Mangieri Antonio, Latib Azeem. Haemodynamic performance and clinical outcomes of transcatheter aortic valve replacement with the self-expanding ACURATE neo2.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2022. DOI: 10.4244/eij-d-22-00289.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.