DOI 10.1016/j.jacc.2022.05.005

Медицинская библиотека · Journal of the American College of Cardiology · 2022

Balloon- vs Self-Expanding Valve Systems for Failed Small Surgical Aortic Valve Bioprostheses.

Rodés-Cabau Josep, Abbas Amr E, Serra Vicenç, Vilalta Victoria, Nombela-Franco Luis, Regueiro Ander, Al-Azizi Karim M, Iskander Ayman, Conradi Lenard, Forcillo Jessica, Lilly Scott, Calabuig Alvaro, Fernandez-Nofrerias Eduard, Mohammadi Siamak, Panagides Vassili, Pelletier-Beaumont Emilie, Pibarot Philippe

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Data comparing valve systems in the valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) field have been obtained from retrospective studies.

OBJECTIVES

The purpose of this study was to compare the hemodynamic results between the balloon-expandable valve (BEV) SAPIEN (3/ULTRA, Edwards Lifesciences) and self-expanding valve (SEV) Evolut (R/PRO/PRO+, Medtronic) in ViV-TAVR.

METHODS

Patients with a failed small (≤23 mm) surgical valve were randomized to receive a BEV or an SEV. The primary endpoint was valve hemodynamics (maximal/mean residual gradients, severe prosthesis patient mismatch [PPM], or moderate-severe aortic regurgitation) at 30 days as evaluated by Doppler echocardiography.

RESULTS

A total of 102 patients were randomized, and of these, 98 patients finally underwent a ViV-TAVR procedure (BEV: n = 46, SEV: n = 52). The procedure was successful in all cases, with no differences in clinical outcomes at 30 days between groups (no death or stroke events). Patients in the SEV group exhibited lower mean and maximal transvalvular gradient values (15 ± 8 mm Hg vs 23 ± 8 mm Hg; P ˂ 0.001; 28 ± 16 mm Hg vs 40 ± 13 mm Hg, P ˂ 0.001), and a tendency toward a lower rate of severe PPM (44% vs 64%; P = 0.07). There were no cases of moderate-severe aortic regurgitation. In total, 55 consecutive patients (SEV: n = 27; BEV: n = 28) underwent invasive valve hemodynamic evaluation during the procedure, with no differences in mean and peak transvalvular gradients between both groups (P = 0.41 and P = 0.70, respectively).

CONCLUSIONS

In patients with small failed aortic bioprostheses, ViV-TAVR with an SEV was associated with improved valve hemodynamics as evaluated by echocardiography. There were no differences between groups in intraprocedural invasive valve hemodynamics and 30-day clinical outcomes (Comparison of the Balloon-Expandable Edwards Valve and Self-Expandable CoreValve Evolut R or Evolut PRO System for the Treatment of Small, Severely Dysfunctional Surgical Aortic Bioprostheses. The 'LYTEN' Trial; NCT03520101).

Для цитирования: Rodés-Cabau Josep, Abbas Amr E, Serra Vicenç, Vilalta Victoria, Nombela-Franco Luis, Regueiro Ander, Al-Azizi Karim M, Iskander Ayman, Conradi Lenard, Forcillo Jessica, Lilly Scott, Calabuig Alvaro, Fernandez-Nofrerias Eduard, Mohammadi Siamak, Panagides Vassili, Pelletier-Beaumont Emilie, Pibarot Philippe. Balloon- vs Self-Expanding Valve Systems for Failed Small Surgical Aortic Valve Bioprostheses.. Journal of the American College of Cardiology. 2022. DOI: 10.1016/j.jacc.2022.05.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.