Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a dead-end.

Noel Antoine, Ploux Sylvain, Bulliard Samuel, Strik Marc, Haeberlin Andreas, Welte Nicolas, Marchand Hugo, Klotz Nicolas, Ritter Philippe, Haïssaguerre Michel, Bordachar Pierre

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantations are rapidly expanding. However, the subcutaneous detection and interpretation of cardiac signals in S-ICDs is much more challenging than by conventional devices. There is a complete paradigm shift in cardiac signal sensing with subcutaneous signal detection, leading in some cases to oversensing with restricted programming options.

OBJECTIVES

The aim of this single-center study was to quantify and describe cases where recurring oversensing made the extraction of the device necessary.

METHODS

Consecutive patients (n = 108) implanted with an S-ICD in our tertiary referral hospital were considered for analysis. Clinical and remote monitoring data were analyzed.

RESULTS

The S-ICD had to be explanted in 6 of 108 implanted patients (5.6%) because of refractory oversensing issues: myopotential oversensing, P- or T-wave oversensing, rate-dependent left bundle branch block aberrancy during exercise with R-wave double counting, and R-wave amplitude decrease after ventricular tachycardia ablation leading to noise detection. Seventeen of 108 patients experienced oversensing (15.7%): 9 patients had at least 1 inappropriate charge without a shock (8.3%), 3 patients had at least 1 inappropriate shock (2.8%), and 5 patients had both episodes (4.6%).

CONCLUSION

So far, cardiologists have had to deal with transvenous ICD lead fractures, but signal oversensing without correcting programming option could be the equivalent weakness of S-ICDs, despite an adequate screening.

Для цитирования: Noel Antoine, Ploux Sylvain, Bulliard Samuel, Strik Marc, Haeberlin Andreas, Welte Nicolas, Marchand Hugo, Klotz Nicolas, Ritter Philippe, Haïssaguerre Michel, Bordachar Pierre. Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a dead-end.. Heart rhythm. 2019. DOI: 10.1016/j.hrthm.2019.07.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.