An automated microaneurysm detector as a tool for identification of diabetic retinopathy in rural optometric practice.

Jelinek Herbert J, Cree Michael J, Worsley David, Luckie Alan, Nixon Peter

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

With the increase in the prevalence of diabetes, rural optometric clinics stand to increase their patient load and assessment of diabetic eye disease. This study aimed to assess whether automated identification of diabetic retinopathy based on the presence of microaneurysms is an effective tool in clinical practice.

METHODS

We analysed 758 fundal images of 385 patients with diabetes attending the clinic obtained using a Canon CR5 with an EOS10 digital camera through a dilated pupil. Five optometrists employed in the clinic assessed the diabetic retinopathy using binocular indirect ophthalmoscopy. The sensitivity and specificity of the automated system used to analyse the retinal fundal images was determined by comparison with optometric and ophthalmologic assessment.

RESULTS

The optometrists achieved 97 per cent sensitivity at 88 per cent specificity with respect to the ophthalmic classification for detecting retinopathy.

CONCLUSION

The automated retinopathy detector achieved 85 per cent sensitivity at 90 per cent specificity at detecting retinopathy. The automated microaneurysm detector has a lower sensitivity compared to the optometrists but meets NHMRC guidelines. It may impact on the efficiency of rural optometric practices by early identification of diabetic retinopathy. Automated assessment can save time and be cost-effective, and provide a history of changes in the retinal fundus and the opportunity for instant patient education using the digital images.

Для цитирования: Jelinek Herbert J, Cree Michael J, Worsley David, Luckie Alan, Nixon Peter. An automated microaneurysm detector as a tool for identification of diabetic retinopathy in rural optometric practice.. Clinical & experimental optometry. 2006. DOI: 10.1111/j.1444-0938.2006.00071.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.