DOI 10.1016/j.jacc.2025.05.011

Медицинская библиотека · Journal of the American College of Cardiology · 2025

Valve Underexpansion and Clinical Outcomes With ACURATE neo2: Findings From the ACURATE IDE Trial.

Makkar Raj R, Chakravarty Tarun, Gupta Aakriti, Soliman Osama, Gnall Eric, Ramana Ravi K, Ramlawi Basel, Diamantouros Pantelis, Potluri Srinivasa, Kleiman Neal S, Samy Sanjay, Rassi Andrew, Yadav Pradeep, Thourani Vinod, Yakubov Steven, Frawley Chris, Patel Dhairya, Kapadia Samir, Chalekian Aaron, Modolo Rodrigo, Sathananthan Janarthanan, Kim Won-Keun, Reardon Michael J

СтатьяРандомизированное исследованиеМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

In the ACURATE IDE (Safety and Effectiveness Study of ACURATE Valve for Transcatheter Aortic Valve Replacement) randomized controlled trial, ACURATE neo2 failed to show noninferiority to commercially available balloon-expandable (SAPIEN 3/3 Ultra) and self-expanding (Evolut R/PRO/PRO+/FX) valves for the primary endpoint of all-cause mortality, stroke, or rehospitalization at 1 year. A retrospective investigation was undertaken to evaluate potential factors contributing to these outcomes.

OBJECTIVES

The goal of this study was to assess the impact of ACURATE neo2 valve expansion on clinical outcomes in the ACURATE IDE trial.

METHODS

Post hoc case review identified angulated (nonparallel) commissure posts in a few implanted ACURATE neo2 valves, indicating valve underexpansion. Procedural angiograms for all ACURATE neo2 valves implanted in the trial's main randomized cohort (n = 752) were inspected by an independent core laboratory. An exploratory analysis was performed to evaluate the association between valve expansion and clinical outcomes.

RESULTS

Of the 624 patients who underwent implantation with the ACURATE neo2 and had evaluable procedural angiograms, 135 (21.6%) had underexpanded valves. Greater aortic valve leaflet and annulus calcification at baseline was independently associated with ACURATE neo2 valve underexpansion (OR: 1.92; 95% CI: 1.27-2.91; P = 0.002). Procedural techniques, including frequency of predilation (100% in both groups) and postdilation (26.7% vs 25.2%; P = 0.72), and balloon sizing did not differ between the underexpanded and expanded valve groups. ACURATE neo2 underexpansion was associated with a higher 1-year rate of death, stroke, or rehospitalization (underexpanded: 18.7%; expanded: 11.8%; P = 0.04), which was confirmed in a multivariable analysis (HR: 1.92; 95% CI: 1.27-2.91; P = 0.002).

CONCLUSIONS

Underexpansion of the ACURATE neo2 valve in the ACURATE IDE study was associated with a higher risk of the composite endpoint of death, stroke, or rehospitalization. Given the post hoc nature of these analyses, the study findings should be considered hypothesis generating. Whether achieving optimal valve expansion of the ACURATE neo2 valve with improvement in device design and procedural iterations will translate into improved clinical outcomes remains to be studied.

Для цитирования: Makkar Raj R, Chakravarty Tarun, Gupta Aakriti, Soliman Osama, Gnall Eric, Ramana Ravi K, Ramlawi Basel, Diamantouros Pantelis, Potluri Srinivasa, Kleiman Neal S, Samy Sanjay, Rassi Andrew, Yadav Pradeep, Thourani Vinod, Yakubov Steven, Frawley Chris, Patel Dhairya, Kapadia Samir, Chalekian Aaron, Modolo Rodrigo, Sathananthan Janarthanan, Kim Won-Keun, Reardon Michael J. Valve Underexpansion and Clinical Outcomes With ACURATE neo2: Findings From the ACURATE IDE Trial.. Journal of the American College of Cardiology. 2025. DOI: 10.1016/j.jacc.2025.05.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.