MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0262159

Медицинская библиотека · PloS one · 2021

The evaluation of the utility of the GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocol.

Naito Asami, Kiyasu Yoshihiko, Akashi Yusaku, Sugiyama Akio, Michibuchi Masashi, Takeuchi Yuto, Notake Shigeyuki, Nakamura Koji, Ishikawa Hiroichi, Suzuki Hiromichi

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

GENECUBE® is a rapid molecular identification system, and previous studies demonstrated that GENECUBE® HQ SARS-CoV-2 showed excellent analytical performance for the detection of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) with nasopharyngeal samples. However, other respiratory samples have not been evaluated.

METHODS

This prospective comparison between GENECUBE® HQ SARS-CoV-2 and reference real-time reverse transcriptase polymerase chain reaction (RT-PCR) was performed for the detection of SARS-CoV-2 using anterior nasal samples and saliva samples. Additionally, we evaluated a new rapid examination protocol using GENECUBE® HQ SARS-CoV-2 for the detection of SARS-CoV-2 with saliva samples. For the rapid protocol, in the preparation of saliva samples, purification and extraction processes were adjusted, and the total process time was shortened to approximately 35 minutes.

RESULTS

For 359 anterior nasal samples, the total-, positive-, and negative concordance of the two assays was 99.7% (358/359), 98.1% (51/52), and 100% (307/307), respectively. For saliva samples, the total-, positive-, and negative concordance of the two assays was 99.6% (239/240), 100% (56/56), and 99.5% (183/184), respectively. With the new protocol, total-, positive-, and negative concordance of the two assays was 98.8% (237/240), 100% (56/56), and 98.4% (181/184), respectively. In all discordance cases, SARS-CoV-2 was detected by additional molecular examinations.

CONCLUSION

GENECUBE® HQ SARS-CoV-2 provided high analytical performance for the detection of SARS-CoV-2 in anterior nasal samples and saliva samples.

Для цитирования: Naito Asami, Kiyasu Yoshihiko, Akashi Yusaku, Sugiyama Akio, Michibuchi Masashi, Takeuchi Yuto, Notake Shigeyuki, Nakamura Koji, Ishikawa Hiroichi, Suzuki Hiromichi. The evaluation of the utility of the GENECUBE HQ SARS-CoV-2 for anterior nasal samples and saliva samples with a new rapid examination protocol.. PloS one. 2021. DOI: 10.1371/journal.pone.0262159.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.