MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s12149-024-01985-7

Медицинская библиотека · Annals of nuclear medicine · 2024

Estimation of liver standardized uptake value in F18-FDG PET/CT scanning: impact of different malignancies, blood glucose level, body weight normalization, and imaging systems.

Abd-Elkader Mohamed S, Elmaghraby Sherif M, Abdel-Mohsen Mohamed A, Khalil Magdy M

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The aim of this work was to investigate homogeneity and stability of liver SUV in terms of different malignancies considering different body normalization schemes and blood glucose concentrations as well as PET/CT imaging systems.

METHODS

The study included 207 patients with four different types of cancers namely breast, lymphoma, lung, and bone-metastasis. Data acquisition was performed with GE Discovery IQ, Biograph mCT, uMI 550, and Ingenuity TF64 after a single intravenous injection of 194 ± 67.5 MBq of 18F-FDG.

RESULTS

In body weight normalization, SUVmax and SUVmean in bone-mets as well as SUVmean in lung patients were not statistically different among scanners especially for data corrected for glucose levels (p = 0.062, 0.121, and 0.150, respectively). In SUVlbm derived from lung patients, there was no significant differences in Philips in comparison to GE and Siemens (both, p > 0.05) for data corrected and not corrected for glucose levels. In SUVbsa, the only non-significant difference revealed among scanners was in the measurements of SUVmean obtained from lung and bone-mets (p = 0.107 and 0.114) both corrected for glucose levels. In SUVbmi, SUVmean of lung and bone-mets as well as SUVmax of bone-mets showed a non-significant differences among the four different scanning systems (p = 0.303, 0.091, and 0.222, respectively) for data corrected for glucose levels.

CONCLUSION

Liver glucose correction needs further investigations in individual tumors but could be potentially affected by whether measurements are made on SUVmean versus SUVmax, body weight normalization, as well as the imaging system. As such, selection of normalization to body weight method should be carefully selected before clinical adoption and clinically adopted and body surface area would provide the highest correlation. As such, normalization of body weight should be carefully made before clinical adoption. SUVmean proves to be useful and stable metric when liver is corrected for blood glucose levels.

Для цитирования: Abd-Elkader Mohamed S, Elmaghraby Sherif M, Abdel-Mohsen Mohamed A, Khalil Magdy M. Estimation of liver standardized uptake value in F18-FDG PET/CT scanning: impact of different malignancies, blood glucose level, body weight normalization, and imaging systems.. Annals of nuclear medicine. 2024. DOI: 10.1007/s12149-024-01985-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.