Diagnostic accuracy of the Novel 29 MHz micro-ultrasound "ExactVuTM" for the detection of clinically significant prostate cancer: A prospective single institutional study. A step forward in the diagnosis of prostate cancer.

Chessa Francesco, Schiavina Riccardo, Ercolino Amelio, Gaudiano Caterina, Giusti Davide, Bianchi Lorenzo, Pultrone Cristian, Marcelli Emanuela, Distefano Concetta, Lodigiani Luca, Brunocilla Eugenio

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION AND

OBJECTIVE

ExactVuTM is a real-time micro-ultrasound system which provides, according to the Prostate Risk Identification Using Micro-Ultrasound protocol (PRI-MUS), a 300% higher resolution compared to conventional transrectal ultrasound. To evaluate the performance of ExactVuTM in the detection of Clinically significant Prostate Cancer (CsPCa).

MATERIALS AND

METHODS

Patients with Prostate Cancer diagnosed at fusion biopsy were imaged with ExactVuTM. CsPCa was defined as any Gleason Score ≥ 3+4. ExactVuTM examination was considered as positive when PRI-MUS score was ≥ 3. PRI-MUS scoring system was considered as correct when the fusion biopsy was positive for CsPCa. A transrectal fusion biopsy- proven CsPCa was considered as a gold standard. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and area under the receiver operator characteristic (ROC) curve (AUC) were calculated.

RESULTS

57 patients out of 68 (84%) had a csPCa. PRI-MUS score was correctly assessed in 68% of cases. Regarding the detection of CsPCa, ExactVuTM 's sensitivity, specificity, PPV, and NPV was 68%, 73%, 93%, and 31%, respectively and the AUC was 0.7 (95% CI 0.5-0-8). For detecting CsPCa in the transition/ anterior zone the sensitivity, specificity, PPV, and NPV was 45%, 66%, 83% and 25% respectively ant the AUC was 0.5 (95% CI 0.2-0.9). Accounting only the CsPCa located in the peripheral zone, sensitivity, specificity, PPV, and NPV raised up to 74%, 75%, 94%, 33%, respectively with AUC 0.75 (95% CI 0.5-0-9).

CONCLUSIONS

ExactVuTM provides high resolution of the prostatic peripheral zone and could represent a step forward in the detection of CsPCa as a triage tool. Further studies are needed to confirm these promising results.

Для цитирования: Chessa Francesco, Schiavina Riccardo, Ercolino Amelio, Gaudiano Caterina, Giusti Davide, Bianchi Lorenzo, Pultrone Cristian, Marcelli Emanuela, Distefano Concetta, Lodigiani Luca, Brunocilla Eugenio. Diagnostic accuracy of the Novel 29 MHz micro-ultrasound "ExactVuTM" for the detection of clinically significant prostate cancer: A prospective single institutional study. A step forward in the diagnosis of prostate cancer.. Archivio italiano di urologia, andrologia : organo ufficiale [di] Societa italiana di ecografia urologica e nefrologica. 2021. DOI: 10.4081/aiua.2021.2.132.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.