Prospective comparison of ultrasound biomicroscopy and anterior segment optical coherence tomography for evaluation of anterior chamber dimensions in European eyes with primary angle closure.
Аннотация
PURPOSE: To compare the accuracy in measurement of the anterior chamber (AC) angle by anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in European patients with suspected primary angle closure (PACS), primary angle closure (PAC), or primary angle-closure glaucoma (PACG).
DESIGN
Cross-sectional study.
METHODS
In all, 55 eyes of 33 consecutive patients presenting with PACS, PAC, or PACG were examined with AS-OCT, followed by UBM. The trabecular-iris angle (TIA) was measured in all four quadrants. The angle-opening distance (AOD) was measured at 500 microm from the scleral spur. The Bland-Altman method was used for assessing agreement between the two methods.
RESULTS
The mean (+/-SD) superior TIA was 19.3+/-15.8 degrees in AS-OCT and 15.7+/-15.0 degrees in UBM (P=0.50) and inferior TIA was 17.9+/-12.9 degrees (AS-OCT) and 16.7+/-14.1 degrees (UBM) (P=0.71). The superior AOD(500) was 0.17+/-0.16 mm in UBM and 0.21+/-0.16 mm in AS-OCT (P=0.06). Bland-Altman analysis showed a mean SD of+/-9.4 degrees for superior and inferior TIA and a mean SD of +/-0.10 mm for superior and inferior AOD(500).
CONCLUSIONS
This comparative study shows that AS-OCT measurements are significantly correlated with UBM measurements but show poor agreement with each other. We do not believe that AS-OCT can replace UBM for the quantitative assessment of the AC angle.