Pacemaker-mediated tachycardia in a biventricular pacing system.

Van Gelder B M, Bracke F A, Meijer A

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 63-year-old man with chronic atrial fibrillation and heart failure had a biventricular pacing system implanted. The pulse generator was a standard DDDR pacemaker, using the atrial channel for the right ventricular lead and the ventricular channel for the left ventricular lead. During final adjustment of the pacing parameters, a pacemaker tachycardia triggered by T wave oversensing from the right ventricular lead was recorded.