Evaluating keratometry and corneal astigmatism data from biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism.

Langenbucher Achim, Taroni Leonardo, Coutinho Catarina P, Cayless Alan, Szentmáry Nóra, Hoffmann Peter, Wendelstein Jascha, Savini Giacomo

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

To compare results from different corneal astigmatism measurement instruments; to reconstruct corneal astigmatism from the postimplantation spectacle refraction and toric intraocular lens (IOL) power; and to derive models for mapping measured corneal astigmatism to reconstructed corneal astigmatism.

METHODS

Retrospective single centre study involving 150 eyes treated with a toric IOL (Alcon SN6AT, DFT or TFNT). Measurements included IOLMaster 700 keratometry (IOLMK) and total keratometry (IOLMTK), Pentacam keratometry (PK) and total corneal refractive power in 3 and 4 mm zones (PTCRP3 and PTCRP4), and Aladdin keratometry (AK). Regression-based models mapping the measured C0 and C45 components (Alpin's method) to reconstructed corneal astigmatism were derived.

RESULTS

Mean C0 components were 0.50/0.59/0.51 dioptres (D) for IOLMK/PK/AK; 0.2/0.26/0.31 D for IOLMTK/PTCRP3/PTCRP4; and 0.26 D for reconstructed corneal astigmatism. All corresponding C45 components ranged around 0. The prediction models had main diagonal elements lower than 1 with some crosstalk between C0 and C45 (nonzero off-diagonal elements). Root-mean-squared residuals were 0.44/0.45/0.48/0.51/0.50/0.47 D for IOLMK/IOLMTK/PK/PTCRP3/PTCRP4/AK.

CONCLUSIONS

Results from the different modalities are not consistent. On average IOLMTK/PTCRP3/PTCRP4 match reconstructed corneal astigmatism, whereas IOLMK/PK/AK show systematic C0 offsets of around 0.25 D. IOLMTK/PTCRP3/PTCRP4. Prediction models can reduce but not fully eliminate residual astigmatism after toric IOL implantation.

Для цитирования: Langenbucher Achim, Taroni Leonardo, Coutinho Catarina P, Cayless Alan, Szentmáry Nóra, Hoffmann Peter, Wendelstein Jascha, Savini Giacomo. Evaluating keratometry and corneal astigmatism data from biometers and anterior segment tomographers and mapping to reconstructed corneal astigmatism.. Clinical & experimental ophthalmology. 2024. DOI: 10.1111/ceo.14387.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.