DOI 10.3389/fmed.2024.1363286

Медицинская библиотека · Frontiers in medicine · 2024

Accuracy of toric intraocular lens power calculation depending on different keratometry values using a novel network based software platform.

Ramsauer Michaela, Luft Nikolaus, Vounotrypidis Efstathios, Priglinger Siegfried G, Mayer Wolfgang J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare different corneal keratometry readings (swept-source-OCT-assisted biometry and Scheimpflug imaging) with a novel software platform for calculation of toric intraocular lenses.

SETTING

Department of Ophthalmology, Ludwig-Maximilians-University, Munich, Germany.

DESIGN

Retrospective, non-randomized, clinical trial.

METHODS

Twenty-three eyes undergoing toric intraocular lens implantation were included. Inclusion criteria were preoperative regular corneal astigmatism of at least 1.00 D, no previous refractive surgery, no ocular surface diseases and no maculopathies. Lens exchange was performed with CALLISTO eye (Zeiss). For each patient, the expected postoperative residual refraction was calculated depending on three different corneal parameters of two different devices: standard K-front (K) and total keratometry (TK) obtained by a swept-source-OCT-assisted biometry system (IOL Master 700, Zeiss) as well as total corneal refractive power (TCRP) obtained by a Scheimpflug device (Pentacam AXL, Oculus). Barrett's formula for toric intraocular lenses was used for all calculations within a novel software platform (EQ workplace, Zeiss FORUM®). Results were statistically compared with postoperative refraction calculated according to the Harris dioptric power matrix.

RESULTS

The standard K values (mean PE 0.02 D ± 0.45 D) and TK values (mean PE 0.09 D ± 0.43 D) of the IOL Master 700 reached similar results (p = 0.96). 78% of eyes in both K and TK groups achieved SE within ±0.5 D of attempted correction and all eyes (100%) were within ±1.0 D of attempted correction in both groups. By contrast, the prediction error in the IOL calculation using the TCRP of the Scheimpflug device was significantly greater (mean PE -0.56 D ± 0.49 D; p = 0.00 vs. standard K and p = 0.00 vs. TK) with adjusted refractive indices. Thirty-nine and Ninety-one percentage of eyes in the TCRP group achieved SE within ±0.5 D (p = 0.008 K vs. TCRP and p = 0.005 TK vs. TCRP) and ± 1.0 D (p = 0.14 vs. TCRP) of attempted correction, respectively.

CONCLUSION

All three corneal parameters (standard K, TK, TCRP) performed well in calculating toric IOLs. The most accurate refractive outcomes in toric IOL implantation were achieved by IOL calculations based on swept-source-OCT-assisted biometry. The SS-OCT-based K-front and TK values achieve comparable results in the calculation of toric IOLs.

Для цитирования: Ramsauer Michaela, Luft Nikolaus, Vounotrypidis Efstathios, Priglinger Siegfried G, Mayer Wolfgang J. Accuracy of toric intraocular lens power calculation depending on different keratometry values using a novel network based software platform.. Frontiers in medicine. 2024. DOI: 10.3389/fmed.2024.1363286.