Comparison of corneal topography maps obtained using the IOLMaster 700® and the Anterion® in candidates for toric IOL implantation.

Chudzinski Roman, Levron Antoine, Agard Emilie, Douma Ikrame, Billant Jérémy, Dot Corinne

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare central corneal topography (CT) obtained using the IOLMaster 700® biometer to corneal topography obtained using a Swept-Source OCT-based predicated topographer (PT), in candidates for toric intraocular lens (IOL) implantation.

METHODS

A retrospective comparative study was conducted in consecutive patients undergoing a routine cataract surgery assessment with significant astigmatism on keratometry. Each patient was examined using both the IOLMaster 700® (Carl Zeiss Meditec, Jena, Germany) and the Anterion® (Heidelberg Engineering, Heidelberg, Germany) for routine preoperative measurements. The corneal axial anterior power map obtained with each device was then anonymized and analysed independently by two ophthalmologists using a reading grid. The reading grid assessed the usual parameters describing astigmatism and evaluated if a toric IOL was indicated or a second topography examination was needed to confirm the indication.

RESULTS

In total, 169 eyes of 120 patients were included. The inter-examination agreement for the astigmatism description ranged from 56 to 85% depending on the reader and parameter. The decision to implant a toric IOL based on the axial map was the same in 59-60% of cases depending on the examiner. A second examination was needed in 18-25% and 8-14% of cases after CT and PT, respectively. The IOLMaster 700® central anterior axial map allowed toric IOL implantation in 58-70% of cases with no need for second corneal examination.

CONCLUSION

The agreement between the anterior axial maps obtained using both devices was good. However, in about a quarter of the cases, dedicated topography had to be performed to confirm the surgical indication.

Для цитирования: Chudzinski Roman, Levron Antoine, Agard Emilie, Douma Ikrame, Billant Jérémy, Dot Corinne. Comparison of corneal topography maps obtained using the IOLMaster 700® and the Anterion® in candidates for toric IOL implantation.. Eye (London, England). 2024. DOI: 10.1038/s41433-024-03094-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.