MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/11206721241245747

Медицинская библиотека · European journal of ophthalmology · 2024

Meridional analysis for calculation of the toric power of phakic IOLs.

Buonsanti Dante, Coutinho Catarina P, Hoffer Kenneth J, Savini Giacomo

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To present a reproducible method to calculate the toricity needed at the intraocular lens (IOL) plane with toric phakic IOLs (ICL, Staar Surgical) and compare its results with those obtained with the online calculator provided by the manufacturer.

DESIGN

Retrospective case series.

SETTING

Private practice, Buenos Aires, Argentina.

METHODS

The formula originally described by Holladay to calculate the IOL power in phakic eyes was used to calculate the required spherical power along the less refractive meridian and along the more refractive meridian. Meridional analysis was applied to calculate the required toricity at the IOL plane and the surgically induced corneal astigmatism was incorporated into the calculations. The refractive cylinder predicted by this method and by the online calculator of the manufacturer were compared to the postoperative refractive cylinder by means of vector analysis. The possible changes in the ratio of toricity in patients with different amounts of astigmatism and anterior chamber depth are assessed in a theoretical section.

RESULTS

In 35 eyes, the measured mean postoperative refractive cylinder was 0.09 D @ 99°, the mean predicted postoperative refractive astigmatism was 0.04 D @ 102° according to the manufacturer's online calculator and 0.09 D @100° according to our method. With both methods, 91.43% of eyes had an absolute cylinder prediction error within ±0.50 diopters.

CONCLUSIONS

The method described in this article to calculate the toricity of phakic IOLs has a refractive accuracy similar to that of the original calculator developed by the manufacturer.

Для цитирования: Buonsanti Dante, Coutinho Catarina P, Hoffer Kenneth J, Savini Giacomo. Meridional analysis for calculation of the toric power of phakic IOLs.. European journal of ophthalmology. 2024. DOI: 10.1177/11206721241245747.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.