Clinical Results of a Trifocal Toric Intraocular Lens Using the Holladay Total Surgically Induced Astigmatism Formula for Correcting Low Corneal Astigmatism in Japanese Patients.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bissen-Miyajima Hiroko, Ota Yuka, Yaguchi Saori, Nakamura Kunihiko, Sasaki Noriyuki
Аннотация
PURPOSE: To evaluate the effectiveness and safety of the AcrySof IQ PanOptix toric intraocular lens (IOL) with cylinder power of 1.0 D (TFNT20) in a Japanese population with low corneal astigmatism and compare with historical control data for nontoric IOLs.
SETTING: Tokyo Dental College Suidobashi Hospital, Tokyo, Japan.
DESIGN: Prospective, single-center study.
METHODS: Patients ≥20 years old received TFNT20 IOL in at least 1 eye based on Alcon Toric calculator (Holladay Total surgically induced astigmatism). Effectiveness endpoints included the percentage of eyes with refractive cylinder ≤0.25 D at 30-60 days after surgery, which was compared with a historical control threshold rate of 29.2% for nontoric IOLs and refractive cylinder ≤0.50 D. Monocular uncorrected distance visual acuity (UDVA; 5 m), uncorrected intermediate visual acuity (UIVA; 60 cm), uncorrected near visual acuity (UNVA; 40 cm), and adverse events were evaluated.
RESULTS: Of 41 eyes implanted with TFNT20 IOLs, 37 eyes (90%) achieved refractive cylinder ≤0.25 D at 30-60 days after surgery, demonstrating the superiority of TFNT20 compared with historical data (P<0.0001). Refractive cylinder of ≤0.50 D was achieved by 41 eyes (100%). At 30-60 days, mean ± SD monocular CDVA was -0.15 ± 0.07 logMAR, UDVA was -0.09 ± 0.09 logMAR, UIVA was -0.00 ± 0.07 logMAR, and UNVA was 0.03 ± 0.07 logMAR. Six eyes (15%) had elevated postoperative intraocular pressure, which returned to normal and was not device-related.
CONCLUSION: TFNT20 IOLs successfully reduced postoperative refractive cylinder and provided good distance, intermediate, and near uncorrected VAs in Japanese patients with low corneal astigmatism.
SETTING: Tokyo Dental College Suidobashi Hospital, Tokyo, Japan.
DESIGN: Prospective, single-center study.
METHODS: Patients ≥20 years old received TFNT20 IOL in at least 1 eye based on Alcon Toric calculator (Holladay Total surgically induced astigmatism). Effectiveness endpoints included the percentage of eyes with refractive cylinder ≤0.25 D at 30-60 days after surgery, which was compared with a historical control threshold rate of 29.2% for nontoric IOLs and refractive cylinder ≤0.50 D. Monocular uncorrected distance visual acuity (UDVA; 5 m), uncorrected intermediate visual acuity (UIVA; 60 cm), uncorrected near visual acuity (UNVA; 40 cm), and adverse events were evaluated.
RESULTS: Of 41 eyes implanted with TFNT20 IOLs, 37 eyes (90%) achieved refractive cylinder ≤0.25 D at 30-60 days after surgery, demonstrating the superiority of TFNT20 compared with historical data (P<0.0001). Refractive cylinder of ≤0.50 D was achieved by 41 eyes (100%). At 30-60 days, mean ± SD monocular CDVA was -0.15 ± 0.07 logMAR, UDVA was -0.09 ± 0.09 logMAR, UIVA was -0.00 ± 0.07 logMAR, and UNVA was 0.03 ± 0.07 logMAR. Six eyes (15%) had elevated postoperative intraocular pressure, which returned to normal and was not device-related.
CONCLUSION: TFNT20 IOLs successfully reduced postoperative refractive cylinder and provided good distance, intermediate, and near uncorrected VAs in Japanese patients with low corneal astigmatism.