DOI 10.1097/j.jcrs.0000000000001429
Clinical outcomes and rotational stability after implantation of a monofocal toric intraocular lens with textured haptics in normal vs high axial lengths.
Аннотация
PURPOSE: To compare the clinical outcomes and rotational stability after implantation of a toric intraocular lens (IOL) with textured haptics in eyes with normal vs high axial lengths (ALs).
SETTING
Nethradhama Superspeciality Eye Hospital, Bangalore, India.
DESIGN
2-arm, retrospective comparative study.
METHODS
This retrospective study included 114 eyes of 114 patients who underwent femtolaser cataract surgery followed by implantation of the HOYA Vivinex Toric monofocal IOL (Model XY1A-SP), of which 62 and 52 eyes belonged to normal (≤23.9 mm) and high (≥24 mm) AL groups, respectively. 1 week and 3 months postoperatively, clinical outcomes and rotational stability of the toric IOL was evaluated.
RESULTS
3 months postoperatively, % eyes achieving refractive astigmatism accuracy within ≤0.50 diopter, was 100% (n = 62) in the normal vs 94% (n = 49) in the high AL group. All eyes that is, 100% (n = 62) in the normal and 96.15% (n = 50) eyes in the high myopia group were <5 degrees of the intended axis. The mean change in postoperative rotation from 1 week to 3 months was 0.28 ± 0.09 degrees in the normal, and 0.30 ± 1.11 degrees in the high AL group ( P = .80). No significant correlation was observed between AL and white-to-white diameter with 1-week postoperative rotation values. No eye required repositioning of toric IOL for significant misalignment.
CONCLUSIONS
No significant differences were observed for clinical outcomes and postoperative rotational stability between eyes with normal and high ALs, suggesting excellent rotational stability of the Vivinex Toric IOL with textured haptics in all eyes, irrespective of the preoperative AL measurements.