MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1155/2012/376384

Медицинская библиотека · Case reports in transplantation · 2012

Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation.

Schubert Sarah A, Soleimani Behzad, Pae Walter E

СтатьяПеревод пока не подготовлен

Аннотация

We report a case of severe hemolysis and pulmonary valve insufficiency (PI) following right ventricular support using a paracorporeal pneumatic pump (Abiomed, Danvers, MA, USA). We speculate that the high velocity jet of blood emanating from the outflow cannula caused turbulence above the pulmonary valve, leading to PI and hemolysis. Despite the growing number of implanted ventricular assist devices, we could find no report in the literature describing pulmonary valve insufficiency secondary to right ventricular assist device (RVAD) placement. Fortunately, in this case, right ventricular function recovered sufficiently after seven days of support, allowing explantation of the device and resolution of PI and hemolysis.

Для цитирования: Schubert Sarah A, Soleimani Behzad, Pae Walter E. Hemolysis and Pulmonary Insufficiency following Right Ventricular Assist Device Implantation.. Case reports in transplantation. 2012. DOI: 10.1155/2012/376384.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.