DOI 10.3928/1081597x-20141021-09
Initial experience with a new refractive rotationally asymmetric multifocal intraocular lens.
Аннотация
PURPOSE: To assess efficacy, safety, predictability, and patient satisfaction after refractive lens exchange with a new refractive rotational asymmetric multifocal intraocular lens.
METHODS
One hundred six eyes of 53 patients after bilateral refractive lens exchange with the SBL-3 lens (Lenstec, Inc., Christ Church, Barbados) implantation were evaluated. The mean preoperative refractive sphere was +1.06 ± 2.63 diopters (D) (range: -8.25 to +5.00 D) and the mean refractive cylinder was -0.51 ± 0.46 D (range: -2.00 to 0.00 D). Monocular and binocular uncorrected and corrected distance visual acuity, uncorrected and distance-corrected intermediate visual acuity, uncorrected and distance-corrected near visual acuity, defocus curve, and patient satisfaction were evaluated 3 months postoperatively.
RESULTS
At 3 months, 84.9% (90 eyes) were within ±0.50 D of emmetropia. The mean postoperative uncorrected distance visual acuity was -0.03 ± 0.09 logMAR (6/6 Snellen) monocularly and -0.08 ± 0.08 logMAR (6/4.8 Snellen) binocularly. The mean monocular and binocular uncorrected near visual acuity were 0.12 ± 0.12 and 0.08 ± 0.10 logMAR (6/7.5 Snellen), respectively. Defocus curve showed a slight drop off for vergences equivalent to intermediate vision. Although some night vision phenomena were reported, overall satisfaction was high. No intraoperative or postoperative complications occurred in this study.
CONCLUSIONS
The new refractive rotationally asymmetric intraocular lens provided good range of vision for near, intermediate, and distance. Long-term follow-ups are necessary to evaluate the performance of this intraocular lens.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.