Removability of a small aperture intracorneal inlay for presbyopia correction.
Аннотация
PURPOSE: To evaluate the safety of the corneal inlay removal procedure and the reversibility of visual acuities, corneal topography, and corneal biomicroscopy changes in a series of cases.
METHODS
Ten cases implanted with one of three versions of the AcuFocus Kamra Inlay (ACI 7000, 7000T, and 7000PDT; AcuFocus, Inc., Irvine, CA) were followed for a minimum of 6 months after corneal inlay removal.
RESULTS
The reason for removal was related to subjective dissatisfaction with visual symptoms (8 of 10 patients) such as night glare, photophobia, starburst, blurry vision, and halos. One case of removal was related to inadvertent thin flap and the final case was related to insufficient near vision. Mean uncorrected distance visual acuity (UDVA) and uncorrected near visual acuity (UNVA) was 0 ± 0.1 logMAR (Snellen 20/20) and 0.5 ± 0.2 logMAR (Snellen 20/40), respectively, preoperatively and 0.1 ± 0.1 logMAR (Snellen 20/25) and 0.5 ± 0.1 logMAR (Snellen 20/63), respectively, 6 months after corneal inlay removal. Mean corrected distance visual acuity (CDVA) and corrected near visual acuity (CNVA) was 0 ± 0.1 logMAR (Snellen 20/20) and 0 ± 0.1 logMAR (Snellen 20/20), respectively, preoperatively and 0 ± 0.1 logMAR (Snellen 20/20) and 0.1 ± 0.1 logMAR (Snellen 20/25), respectively, 6 months after corneal inlay removal. Mean root mean square (RMS) higher-order aberration (HOA) was 0.50 ± 0.12 (range: 0.30 to 0.70) preoperatively and 0.69 ± 0.14 (range: 0.48 to 0.95) 6 months after corneal inlay removal (P < .8). Weak positive correlation was found between Δt Implant-Removal (Δt I-R), RMS spherical, coma, and HOA at 6 months (Δt I-R vs RMS spherical was r = 0.2, r(2) = 0.5, P < .7; Δt I-R vs RMS coma was r = 0.8, r(2) = 0.6, P < .3; and Δt I-R vs HOA r = 0.8; r(2) = 0.6, P < .9).
CONCLUSION
This study suggests that after removal of the corneal inlay, corneal topography and corneal aberrometry are not permanently affected. In more than 60% of patients, CNVA, CDVA, UNVA, and UDVA were similar to the preoperative value.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.