Visual Strehl performance of IOL designs with extended depth of focus.

Gallego Augusto Arias, Bará Salvador, Jaroszewicz Zbigniew, Kolodziejczyk Andrzej

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The use of monofocal, bifocal, or multifocal intraocular lenses (IOLs) is a common option to restore the refractive power of the eye in aphakic patients after cataract surgery. In this article, we study the predicted performance of two new designs of IOL, both with extended depth of focus: the quartic axicon and the light sword optical element (LSOE). These elements provide continuous focal segments spanning the range of dioptric power required for presbyopia compensation.

METHODS

The performance analysis is based on the visual Strehl ratio (SR) computed in the spatial frequency domain, a neuro-optical quality metric that takes into account the effects of both the optical transfer function and the neural contrast sensitivity function. Furthermore, the classical SR and compound modulation transfer function were calculated. Some conventional transmittances of commercially available IOLs are also analyzed.

RESULTS

The LSOE design has a more homogeneous behavior than other available solutions, providing a more uniform image quality over a significant fraction of the required addition range.

CONCLUSIONS

The angular average of the visual SR computed in the spatial frequency domain and compound modulation transfer function indicate that both designs of extended depth of focus elements provide better optical quality on the whole addition range, except at the end points, than the discrete focus designs. The LSOE performed better than the quartic axicon in terms of uniformity of image quality within the same range.

Для цитирования: Gallego Augusto Arias, Bará Salvador, Jaroszewicz Zbigniew, Kolodziejczyk Andrzej. Visual Strehl performance of IOL designs with extended depth of focus.. Optometry and vision science : official publication of the American Academy of Optometry. 2012. DOI: 10.1097/opx.0b013e3182775e1a.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.