DOI 10.1016/s0886-3350(00)00409-0
Potential acuity meter accuracy in cataract patients.
Аннотация
PURPOSE: To determine potential acuity meter (PAM) accuracy in patients with cataract as the only ocular disease.
SETTING
Hospital de Clínicas de Porto Alegre, Brazil.
METHODS
This prospective study comprised 74 eyes with cataract. Patients with other ocular pathology diagnosed before or after surgery, as well as those with complicated surgeries, were excluded. All patients had PAM evaluation before surgery; these results were compared to the final postoperative visual acuity. Patients were divided into 4 cataract groups based on the Lens Opacity Classification System II: Group 1, early cataracts (nuclear opalescence [NO] 1 and/or posterior subcapsular cataracts [P] 1 and/or cortical opacity [C] < 2); Group 2, definite cataracts (NO2 and/or C2, with or without P1); Group 3, posterior cataracts (P2 or P3); Group 4, mature cataracts (N3 and/or C > or = 3, with or without P1). The PAM result was considered accurate if it was within 2 lines of the postoperative best spectacle-corrected visual acuity (BSCVA) 3 months after surgery.
RESULTS
All patients had a BSCVA of 20/20. Accuracy of the PAM was 58.3% in Group 1, 50.0% in Group 2, 27.8% in Group 3, and 6.7% in Group 4 (P =.000 017). Patient age did not influence the results.
CONCLUSIONS
The PAM may be used as a complementary tool in evaluating early and nondense cataracts; however, patients should not be excluded as candidates for cataract surgery based solely on PAM results.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.