Digestive diseases and sciences · 2023
Defecatory Function Studies Using the Fecobionics Device Are Repeatable.
СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lo Kar Man, Sun Daming, Chen Ssu-Chi, Leung Wing Wa, Wong Cherry, Mak Tony, Ng Simon, Futaba Kaori, Gregersen Hans
Аннотация
BACKGROUND: Only limited data exist on repeatability of anorectal studies with the established physiological and clinical technologies for assessment of anorectal function. Fecobionics is a new multi-sensor simulated feces that provide data by integrating elements from current tests.
AIMS: To study repeatability of anorectal data obtained with the Fecobionics device.
METHODS: We assessed the database of Fecobionics studies to determine how many repeated studies were done. From a total of 260 Fecobionics studies, 19 subjects with repeated studies using approximately the same protocol and prototype were identified. Key pressure and bending parameters were assessed and the repeatability analyzed using Bland Altman plots. Furthermore, the inter- and intra-individual coefficient of variation (CV) were computed.
RESULTS: Fifteen subjects (5F/10 M) with repeated studies were normal subjects, three were patients with fecal incontinence and one subject suffered from chronic constipation. The main analysis was conducted on the cohort of normal subjects. The bias for 11 parameters were within the confidence interval, whereas two were slightly outside. The interindividual CV was lowest for the bend angle (10.1-10.7) and between 16.3 and 51.6 for the pressure parameters. The intra-individual CVs were approximately half of the inter-individual CVs, spanning from 9.7 to 27.6.
CONCLUSION: All data from normal subjects were within previously defined normality. The Fecobionics data showed acceptable repeatability with bias within the confidence limits for almost all parameters. The intra-individual CV was much lower than the inter-individual CV. Dedicated large-scale studies are warranted to evaluate the influence of age, sex, and disease on repeatability as well as comparing between technologies.
AIMS: To study repeatability of anorectal data obtained with the Fecobionics device.
METHODS: We assessed the database of Fecobionics studies to determine how many repeated studies were done. From a total of 260 Fecobionics studies, 19 subjects with repeated studies using approximately the same protocol and prototype were identified. Key pressure and bending parameters were assessed and the repeatability analyzed using Bland Altman plots. Furthermore, the inter- and intra-individual coefficient of variation (CV) were computed.
RESULTS: Fifteen subjects (5F/10 M) with repeated studies were normal subjects, three were patients with fecal incontinence and one subject suffered from chronic constipation. The main analysis was conducted on the cohort of normal subjects. The bias for 11 parameters were within the confidence interval, whereas two were slightly outside. The interindividual CV was lowest for the bend angle (10.1-10.7) and between 16.3 and 51.6 for the pressure parameters. The intra-individual CVs were approximately half of the inter-individual CVs, spanning from 9.7 to 27.6.
CONCLUSION: All data from normal subjects were within previously defined normality. The Fecobionics data showed acceptable repeatability with bias within the confidence limits for almost all parameters. The intra-individual CV was much lower than the inter-individual CV. Dedicated large-scale studies are warranted to evaluate the influence of age, sex, and disease on repeatability as well as comparing between technologies.