Loss of sensitivity of immunochromatographic test (ICT) for lymphatic filariasis diagnosis in low prevalence settings: consequence in the monitoring and evaluation procedures.

Gounoue-Kamkumo Raceline, Nana-Djeunga Hugues C, Bopda Jean, Akame Julie, Tarini Ann, Kamgno Joseph

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Diagnostic tools for lymphatic filariasis (LF) elimination programs are useful in mapping the distribution of the disease, delineating areas where mass drug administrations (MDA) are required, and determining when to stop MDA. The prevalence and burden of LF have been drastically reduced following mass treatments, and the evaluation of the performance of circulating filarial antigen (CFA)-based assays was acknowledged to be of high interest in areas with low residual LF endemicity rates after multiple rounds of MDA. The objective of this study was therefore to evaluate the immunochromatographic test (ICT) sensitivity in low endemicity settings and, specifically, in individuals with low intensity of lymphatic filariasis infection.

METHODS

To perform this study, calibrated thick blood smears, ICT and Og4C3 enzyme-linked immunosorbent assay (ELISA) were carried out by night to identify Wuchereria bancrofti microfilarial and circulating filarial antigen carriers. A threshold determination assay regarding ICT and ELISA was performed using serial plasma dilutions from individuals with positive microfilarial counts.

RESULTS

All individuals harbouring microfilariae (positive blood films) were detected by ICT and ELISA, but among individuals positive for ELISA, only 35.7 % of them were detected using ICT (Chi square: 4.57; p-value = 0.03), indicating a moderate agreement between both tests (kappa statistics = 0.49). Threshold determination analyses showed that ELISA was still positive at the last plasma dilution with negative ICT result.

CONCLUSIONS

These findings suggest a loss of sensitivity for ICT in low endemicity settings, especially in people exhibiting low levels of circulating filarial antigen, raising serious concern regarding the monitoring and evaluation procedures in the framework of LF elimination program.

Для цитирования: Gounoue-Kamkumo Raceline, Nana-Djeunga Hugues C, Bopda Jean, Akame Julie, Tarini Ann, Kamgno Joseph. Loss of sensitivity of immunochromatographic test (ICT) for lymphatic filariasis diagnosis in low prevalence settings: consequence in the monitoring and evaluation procedures.. BMC infectious diseases. 2015. DOI: 10.1186/s12879-015-1317-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.