MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/bjophthalmol-2020-317124

Медицинская библиотека · The British journal of ophthalmology · 2020

Impact of intraocular lens characteristics on intraocular lens dislocation after cataract surgery.

Mayer-Xanthaki Christoph Fidel, Pregartner Gudrun, Hirnschall Nino, Falb Thomas, Sommer Michael, Findl Oliver, Wedrich Andreas

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To assess the influence of intraocular lens (IOL) characteristics on IOL dislocations after cataract surgery.

METHODS

Patients who underwent cataract surgery at the Department of Ophthalmology Graz, Austria, between 1996 and 2017 were included and medical records were reviewed. Cox proportional-hazard regression models were used to assess the influence of IOL characteristics on IOL dislocation.

RESULTS

From 68 199 eyes out of 46 632 patients (60.2% women, mean age: 73.71 SD±10.82 years), 111 (0.16%) had an in-the-bag (ITB) disloaction and 35 (0.05%) had an out-of-the-bag (OTB) dislocation. The HRs adjusted for predisposing factors significantly associated with a higher risk for an ITB dislocation were 2.35 (95% CI, 1.45 to 3.8) for hydrophilic IOLs, 2.01 for quadripode IOLs (95% CI, 1.04 to 3.86) and 1.61 (95% CI, 1.04 to 2.48) for haptic angulation. A lower risk was observed for three-piece IOLs (HR=0.58, 95% CI, 0.34 to 0.98) and larger overall IOL diameter (HR=0.79, 95% CI, 0.66 to 0.95). For an OTB dislocation, the HR associated with a higher risk was 18.81 (95% CI, 5.84 to 60.58) for silicone IOLs and 2.12 (95% CI, 0.62 to 7.29) for hydrophilic IOLs. Larger overall IOL diameter (HR 0.40, 95% CI; 0.25 to 0.63) showed a lower risk.

CONCLUSION

Hydrophilic IOLs, quadripode IOLs and haptic angulation were associated with a higher risk for an ITB dislocation, whereas three-piece IOLs and a larger overall diameter were associated with a lower risk. Risk factors for OTB dislocation were silicone IOLs, hydrophilic IOLs and a smaller overall IOL diameter.

Для цитирования: Mayer-Xanthaki Christoph Fidel, Pregartner Gudrun, Hirnschall Nino, Falb Thomas, Sommer Michael, Findl Oliver, Wedrich Andreas. Impact of intraocular lens characteristics on intraocular lens dislocation after cataract surgery.. The British journal of ophthalmology. 2020. DOI: 10.1136/bjophthalmol-2020-317124.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.