Change in optical axial length after cataract surgery: segmental method vs composite method.
Аннотация
PURPOSE: To compare the changes in axial length (AL) obtained after cataract surgery by partial coherence interferometry (PCI), swept-source optical coherence tomography with the composite method (SS-OCT/CM), and SS-OCT with the segmental method (SS-OCT/SM).
SETTING
National Hospital Organization, Tokyo Medical Center, Japan.
DESIGN
Retrospective case series.
METHODS
AL was measured preoperatively and 1 month postoperatively by PCI, SS-OCT/CM, and SS-OCT/SM.
RESULTS
Two hundred sixty-eight eyes in 139 patients who underwent cataract surgery were included. The mean ALs measured by PCI and SS-OCT/CM decreased significantly from 24.14 ± 1.58 to 24.05 ± 1.58 mm (P < .0001) and from 24.15 ± 1.58 to 24.05 ± 1.59 mm (P < .0001), respectively. The mean differences between the preoperative and postoperative AL were 0.083 ± 0.044 mm (PCI) and 0.096 ± 0.045 mm (SS-OCT/CM), respectively; Bland-Altman analysis revealed fixed bias between the preoperative and postoperative values. The mean preoperative and postoperative ALs measured by SS-OCT/SM were 24.12 ± 1.54 and 24.12 ± 1.54 mm, respectively (P = 0.97). The mean difference between the preoperative and postoperative ALs was 0.00 ± 0.03 mm on SS-OCT/SM; Bland-Altman analysis indicated good agreement between these values.
CONCLUSIONS
The difference between preoperative and postoperative ALs is less with SS-OCT/SM than with PCI or SS-OCT/CM. Assuming that AL is not altered by cataract surgery, AL measurement is more accurate by SS-OCT/SM than by PCI or SS-OCT/CM.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.