PloS one · 2019
Comparison of two swept-source optical coherence tomography biometers and a partial coherence interferometer.
Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yang Chan Min, Lim Dong Hui, Kim Hyo Jeong, Chung Tae-Young
Аннотация
PURPOSE: To compare biometry and prediction of postoperative refractive outcomes obtained by two swept-source optical coherence tomography (SS-OCT) biometers (IOLMaster 700 and Argos), and a partial coherence interferometry (IOLMaster ver 5.4).
METHODS: Biometric values were measured using two SS-OCT and PCI device and evaluated against one another. Predictive errors were compared at one month after cataract surgery.
RESULTS: One hundred forty six eyes were considered. Axial length (AXL) measurements were not successful in 3 eyes measured by IOLMaster 700 and Argos devices, and in 17 eyes measured by IOLMaster ver. 5.4 devices. AXL as measured by Argos showed a tendency to be shorter in long eyes with AXL more than 26.0 mm (p < .001) and to be longer in short eyes with AXL less than 22.5 mm (p = .005). Anterior chamber depth as measured by IOLMaster ver. 5.4 was longer than that measured by the other two SS-OCT devices (vs. IOLMaster 700: p = .003; vs. Argos: p = .006). White-to-white diameter measured using Argos was significantly different measurements obtained using two IOLMaster (p < .001, respectively). The mean absolute postoperative prediction errors were 0.41 ± 0.31 diopters (D), 0.42 ± 0.32 D, and 0.35 ± 0.30 D for IOLMaster ver. 5.4, IOLMaster 700, and Argos, respectively.
CONCLUSION: The ocular biometric measurements using three devices showed high agreement. AXL measured by Argos showed a significant difference compared with the measurements from two IOLMaster. ACD was highly correlated between two SS-OCT devices except IOLMaster ver 5.4. LT and CCT values between IOLMaster 700 and Argos were different significantly. SS-OCT devices demonstrated a superior ability to successfully perform measurements compared with PCI device.
METHODS: Biometric values were measured using two SS-OCT and PCI device and evaluated against one another. Predictive errors were compared at one month after cataract surgery.
RESULTS: One hundred forty six eyes were considered. Axial length (AXL) measurements were not successful in 3 eyes measured by IOLMaster 700 and Argos devices, and in 17 eyes measured by IOLMaster ver. 5.4 devices. AXL as measured by Argos showed a tendency to be shorter in long eyes with AXL more than 26.0 mm (p < .001) and to be longer in short eyes with AXL less than 22.5 mm (p = .005). Anterior chamber depth as measured by IOLMaster ver. 5.4 was longer than that measured by the other two SS-OCT devices (vs. IOLMaster 700: p = .003; vs. Argos: p = .006). White-to-white diameter measured using Argos was significantly different measurements obtained using two IOLMaster (p < .001, respectively). The mean absolute postoperative prediction errors were 0.41 ± 0.31 diopters (D), 0.42 ± 0.32 D, and 0.35 ± 0.30 D for IOLMaster ver. 5.4, IOLMaster 700, and Argos, respectively.
CONCLUSION: The ocular biometric measurements using three devices showed high agreement. AXL measured by Argos showed a significant difference compared with the measurements from two IOLMaster. ACD was highly correlated between two SS-OCT devices except IOLMaster ver 5.4. LT and CCT values between IOLMaster 700 and Argos were different significantly. SS-OCT devices demonstrated a superior ability to successfully perform measurements compared with PCI device.