MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2019.03.003

Медицинская библиотека · Journal of cataract and refractive surgery · 2019

Optical behavior of the eye implanted with extreme intraocular lens powers.

Oliveira Renan F, Salerno Liberdade C, Mimouni Michael, Plaza-Puche Ana Belén, Alió Jorge L

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the induced ocular aberrations after cataract surgery with extreme low- powered and high-powered spherical monofocal intraocular lenses (IOLs) using a new pyramidal wavefront sensor aberrometer.

SETTING

Vissum Instituto Oftalmológico, Alicante, Spain.

DESIGN

Prospective observational comparative study.

METHODS

This study included patients who had cataract surgery with implantation of a spherical monofocal IOL. The patients were divided into three groups according to the IOL power. Group 1: +20 diopters (D) to +23 D; Group 2: less than +10 D; and Group 3: +29 D or more. Assessment after 3 months included: uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, contrast sensitivity function test (Ginsburg), corneal aberrometry, and total ocular aberrometry with a pyramidal wavefront sensor aberrometer (Osiris) for 4.0 mm, 4.5 mm, and 5.0 mm pupil diameters.

RESULTS

The study comprised 88 eyes of 56 patients. There were no differences between all groups in terms of refractive error, corneal aberrometry, or total ocular higher-order aberrations at any analyzed pupil diameter. The ocular spherical aberration (SA) was not significantly different between groups with a 4.0 mm pupil. The SA was significantly lower in Group 2 when compared with Group 1 and Group 3 for both 4.5 mm (P = .01 and P = .001, respectively) and 5.0 mm (P = .002 and P = .002, respectively). There was no significant difference in SA between Group 1 (+20 D to 23 D) and Group 3 (≥+29 D) at any analyzed pupil diameter. Higher SA did not correlate to lower contrast sensitivity.

CONCLUSIONS

Despite the theoretical evidence, only a negligible amount of clinically insignificant aberrations were induced by high-powered positive IOLs. Low-powered IOLs did not induce significant levels of SAs.

Для цитирования: Oliveira Renan F, Salerno Liberdade C, Mimouni Michael, Plaza-Puche Ana Belén, Alió Jorge L. Optical behavior of the eye implanted with extreme intraocular lens powers.. Journal of cataract and refractive surgery. 2019. DOI: 10.1016/j.jcrs.2019.03.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.