MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jvim.17170

Медицинская библиотека · Journal of veterinary internal medicine · 2024

Use of prototype bi-nasal prongs for noninvasive ventilation in foals.

Raidal Sharanne L, van Diggelen Michael, Catanchin Chee Sum M, Lehmann Heidi S, Quinn Chris T

СтатьяRandomized Controlled Trial, VeterinaryПеревод пока не подготовлен

Аннотация

BACKGROUND

Noninvasive ventilation (NIV) provides effective respiratory support in foals, but face masks are poorly tolerated and associated with hypercapnia. Bi-nasal prongs might be a more effective device interface in foals.

OBJECTIVES

To compare bi-nasal prongs and masks for NIV in foals with pharmacologically induced respiratory insufficiency.

ANIMALS: Six healthy foals.

METHODS

In a randomized cross-over study, sedated foals received NIV delivered by mask or bi-nasal prongs, with the treatment repeated using the alternative device interface after a 3-day rest period. After periods of spontaneous ventilation through the allocated interface, with and without supplementary O2 (T2-T3), foals were subject to 10-minute treatment periods of NIV at different pressure support (5 or 10 cmH2O) and end-expiratory pressure settings (5 or 10 cmH2O), with and without supplementary O2 (T4-T7). Vital signs, arterial blood gases, spirometry, and gas exchange data were measured in the final 2 minutes of each treatment window.

RESULTS

Bi-nasal prongs were well tolerated and required less manual positioning or monitoring compared to the mask. Partial pressure of carbon dioxide did not increase during NIV with bi-nasal prongs and was lower than observed with masks (mean difference, 8.2 mmHg [95% confidence interval, 4.1-12.2 mmHg] at T6). Oxygenation and respiratory mechanics were improved in all foals and not different between device interfaces.

CONCLUSIONS AND CLINICAL

IMPORTANCE

Nasal prongs were well tolerated, had similar effects on respiratory function, and appeared to ameliorate hypercapnia observed previously during NIV in foals.

Для цитирования: Raidal Sharanne L, van Diggelen Michael, Catanchin Chee Sum M, Lehmann Heidi S, Quinn Chris T. Use of prototype bi-nasal prongs for noninvasive ventilation in foals.. Journal of veterinary internal medicine. 2024. DOI: 10.1111/jvim.17170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.