DOI 10.1097/icl.0000000000001037

Медицинская библиотека · Eye & contact lens · 2023

Optical Quality and Visual Performance With Different Toric Contact Lens Designs.

Koh Shizuka, Maeda Naoyuki, Terao Misaki, Maeda Hiroko, Kosaki Ryo, Kozaki Jun, Nishida Kohji

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To evaluate the optical quality of different toric contact lens (CL) designs and compare their on-eye visual correction performance.

METHODS

Twenty soft CL wearers aged 20 to 39 years were enrolled. Two daily disposable silicone-hydrogel toric CLs were tested: the "Eyelid Stabilized Design" (ESD-CL) and prism-ballast design (PB-CL); a spherical daily disposable silicone hydrogel CL (spherical CL) was used as a control. On-eye performance was compared for corrected distance visual acuity (CDVA), astigmatism, and ocular higher-order aberrations (HOAs); astigmatism and ocular HOAs were measured with a wavefront sensor. The subjective quality of vision, rated for "blurred vision" and "double vision," lens rotation, and fitting were also compared.

RESULTS

The ESD-CLs, PB-CL, and no-CL provided better CDVA than spherical CL ( P <0.05). Compared with spherical CL and no CL, PB-CL and ESD-CLs caused significantly lesser astigmatism ( P <0.05). Coma was significantly lesser with ESD-CLs than that with PB-CL ( P <0.05); total HOAs did not differ among the four conditions. The subjective ratings for blurred and double vision were significantly lower with ESD-CLs than those with spherical CL ( P <0.05).

CONCLUSIONS

Toric CLs provide a better CDVA than spherical CLs. However, differences in coma and subjective symptoms may occur because of the design of toric CLs.

Для цитирования: Koh Shizuka, Maeda Naoyuki, Terao Misaki, Maeda Hiroko, Kosaki Ryo, Kozaki Jun, Nishida Kohji. Optical Quality and Visual Performance With Different Toric Contact Lens Designs.. Eye & contact lens. 2023. DOI: 10.1097/icl.0000000000001037.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.