MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacep.2020.02.003

Медицинская библиотека · JACC. Clinical electrophysiology · 2020

Acute Hemoptysis Following Cryoballoon Pulmonary Vein Isolation: A Multicenter Study.

Vogler Julia, Fink Thomas, Sohns Christian, Sommer Philipp, Pott Alexander, Dahme Tillman, Rottner Laura, Sciacca Vanessa, Sieren Malte Maria, Jacob Fabian, Barkhausen Jörg, Sano Makoto, Eitel Charlotte, Metzner Andreas, Ouyang Feifan, Kuck Karl-Heinz, Tilz Roland Richard, Heeger Christian-Hendrik

Клинический случайСтатьяМногоцентровое исследованиеНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to assess the incidence, procedural characteristics, contributing factors, and clinical outcome of cryoballoon-based pulmonary vein isolation (CB-PVI)-related hemoptysis in a multicenter study.

BACKGROUND

Hemoptysis has been described as a rare complication of CB-PVI. However, the precise mechanism and the etiology of this complication are poorly characterized.

METHODS

Consecutive patients undergoing CB-PVI for paroxysmal or persistent atrial fibrillation at 4 German hospitals were included in this observational analysis.

RESULTS

A total of 4,331 CB-PVI procedures were performed between 2006 and 2019. Fifteen patients (9 men, mean age 68.1 ± 9.8 years) developed acute hemoptysis during or within 24 h after CB-PVI, resulting in a hemoptysis frequency of 0.35%. Hemoptysis occurred in 6 of 720 procedures using the first-generation CB (0.83%) and in 9 of 3,611 procedures using the second-, third-, or fourth-generation CB (0.25%) (p = 0.015). Bronchoscopy was performed in 8 patients and showed bleeding exclusively due to mucosal injury or due to a coagulum at a bronchus adjacent to the ablation site. Hemoptysis resolved spontaneously without any long-term sequelae in all patients, except for a 92-year-old patient who died 13 days after CB-PVI due to pneumonia. No specific endobronchial treatment was necessary.

CONCLUSIONS

Acute hemoptysis after CB-PVI is a rare but potentially life-threatening complication that is usually self-limiting. Direct thermal injury of bronchi adjacent to a pulmonary vein seems to be the most likely mechanism.

Для цитирования: Vogler Julia, Fink Thomas, Sohns Christian, Sommer Philipp, Pott Alexander, Dahme Tillman, Rottner Laura, Sciacca Vanessa, Sieren Malte Maria, Jacob Fabian, Barkhausen Jörg, Sano Makoto, Eitel Charlotte, Metzner Andreas, Ouyang Feifan, Kuck Karl-Heinz, Tilz Roland Richard, Heeger Christian-Hendrik. Acute Hemoptysis Following Cryoballoon Pulmonary Vein Isolation: A Multicenter Study.. JACC. Clinical electrophysiology. 2020. DOI: 10.1016/j.jacep.2020.02.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.