MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijcard.2014.07.241

Медицинская библиотека · International journal of cardiology · 2014

Cerebral microemboli and neurocognitive change after carotid artery stenting with different embolic protection devices.

Akkaya Emre, Vuruskan Ertan, Gul Zeynep Bastug, Yildirim Aydın, Pusuroglu Hamdi, Surgit Ozgur, Kalkan Ali Kemal, Akgul Ozgur, Akgul Gamze Pinar, Gul Mehmet

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Proximal cerebral protection devices have been developed as an alternative to filter protection devices for reducing neurological complications during carotid artery stenting (CAS). The aim of the present study was to evaluate the frequency of silent cerebral embolism after CAS using different cerebral embolic protection devices and the impact of silent cerebral embolism on neurocognitive function.

METHODS

One hundred consecutive patients who underwent CAS were enrolled. The patients were randomized to either proximal balloon occlusion or filter protection. Neurocognitive tests were performed before and six months after CAS. Cerebral embolisms were evaluated with diffusion-weighted magnetic resonance imaging (DW-MRI).

RESULTS

The number and volume of new ischemic lesions found with DW-MRI were higher in the filter protection group than in the proximal balloon occlusion group. According to our definition, nine (21%) patients in the balloon occlusion group and 16 (36%) patients in the filter protection group showed neurocognitive decline, and ten (23%) patients in the balloon occlusion group and four (9%) patients in the filter protection group showed neurocognitive improvement (NS). Regarding the group of patients with new cerebral ischemic lesions on DW-MRI, neurocognitive decline occurred in 14 (31%) of 45 patients with DW-MRI lesions and 11 (26%) of 43 patients without DW-MRI lesions (NS).

CONCLUSION

Neurocognitive outcome after CAS is unpredictable; both neurocognitive decline and improvement can occur. In this study, the proximal balloon occlusion system significantly decreased cerebral microemboli during CAS compared to filter protection. Cerebral microembolism was not found to be associated with neurocognitive decline.

Для цитирования: Akkaya Emre, Vuruskan Ertan, Gul Zeynep Bastug, Yildirim Aydın, Pusuroglu Hamdi, Surgit Ozgur, Kalkan Ali Kemal, Akgul Ozgur, Akgul Gamze Pinar, Gul Mehmet. Cerebral microemboli and neurocognitive change after carotid artery stenting with different embolic protection devices.. International journal of cardiology. 2014. DOI: 10.1016/j.ijcard.2014.07.241.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.