DOI 10.1373/clinchem.2006.070730

Медицинская библиотека · Clinical chemistry · 2006

Performance and clinical utility of a commercial von Willebrand factor collagen binding assay for laboratory diagnosis of von Willebrand disease.

Popov Joanna, Zhukov Olga, Ruden Susan, Zeschmann Terry, Sferruzza Anthony, Sahud Mervyn

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Von Willebrand disease (VWD) diagnosis and classification usually require a combination of nonspecific and VW-factor (VWF)-specific assays. We evaluated the analytical performance of a commercially available collagen-binding assay (CBA) and its usefulness in conjunction with other assays for laboratory diagnosis of VWD.

METHODS

We used a commercial CBA ELISA (Life Technologies) to evaluate 3085 plasma samples. We used standard procedures to perform other assays, including factor VIII activity (FVIII:C), VWF antigen (VWF:Ag), ristocetin cofactor activity, VWF collagen binding capacity (VWF:CB), and VWF multimeric analysis.

RESULTS

CBA intra- and interassay CVs were <6% and <13%, respectively. Reference intervals were 45%-198% for VWF:CB and 0.75-1.32 for the VWF:CB/Ag ratio. Of 3085 samples tested, 235 (8%) had results commonly associated with VWD. Multimer analysis and phenotypic data in 156 samples identified VWD types as: 91 (58%) type 1, 62 (40%) type 2, and 3 (2%) type 3. Of the 91 type 1 samples, proportional decreases in functional activity were seen in 75 samples (82%) according to CBA and in 63 samples (69%) according to the ristocetin cofactor assay. Of the type 2 samples, 10 were further identified as probable type 2A, 26 as probable type 2B, 12 as probable type 2M, and 14 could not be subtyped. VWF:CBA/Ag ratios <0.5 occurred in 83% of VWD type 2A and 2B samples, indicating characteristic functional discordance. Mean (SD) VWF:CB values were significantly higher in individuals without group O blood [113 (45)] than in those with group O blood [83 (32)] (t-test, P = 0.007).

CONCLUSIONS

The commercial CBA assay produces reliable results and is useful for laboratory diagnosis of VWD.

Для цитирования: Popov Joanna, Zhukov Olga, Ruden Susan, Zeschmann Terry, Sferruzza Anthony, Sahud Mervyn. Performance and clinical utility of a commercial von Willebrand factor collagen binding assay for laboratory diagnosis of von Willebrand disease.. Clinical chemistry. 2006. DOI: 10.1373/clinchem.2006.070730.

Предметные рубрики MeSH