Медицинская библиотека · Oftalmologia (Bucharest, Romania : 1990) · 2013

[Different significance in normal subjects and in glaucoma patients tested with Optopol PTS-910, in the galucoma program].

Dascalu Ana Maria, Cherecheanu Alina Popa, Stana Daniela, Serban Dragoş

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: to quantify the inter-test variability (dB) for the Optopol PTS automated perimeter, Glaucoma Fast threshold program.

MATERIALS AND METHOD: A prospective study was performed on 166 glaucomatous patients and a control group of 30 normal subjects, tested by complete ophthalmological exam and automated perimetry (Optopol PTS-910). The visual field was tested weekly for 4 consecutive weeks. The visual field defects were classified according to the Aulhorn-Karmeyer descriptive scale.

RESULTS

For the control group, the medium inter-test variability was of 1.57 +/- 0.24 dB, lower next to fixation and increasing towards the 50 degree isopter. The medium inter-test variability increases along with the perimetric stage :1.57 +/- 0.66 dB for pre-perimetric glaucoma, 2.13 + 1.04 dB for non-specific defects group, 3.23 + 1.01 dB for the stage 1, 3.52 + 2.61 dB, for the stage 2, 3.65 + 1.19dB for the stage 3 and 5.82 +/- 1.67dB for the stage 4. For the cases of preperimetric glaucoma and non-specific defects, a similar profile of variability to the normal subjects can be observed. For the stages 2-4, the profile of the areas with maxim inter-test variability moves towards the relative scotoma and the surrounding area.

CONCLUSIONS

A better description of the inter-test variability and the evolution of this intricate parameter of the retinal light sensitivity is useful for the differential diagnostic between the real change and the "background noise" in early detection of the functional progression in glaucoma.

Для цитирования: Dascalu Ana Maria, Cherecheanu Alina Popa, Stana Daniela, Serban Dragoş. [Different significance in normal subjects and in glaucoma patients tested with Optopol PTS-910, in the galucoma program].. Oftalmologia (Bucharest, Romania : 1990). 2013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.