DOI 10.3343/alm.2019.39.2.176

Медицинская библиотека · Annals of laboratory medicine · 2019

Comparative Evaluation Between the RealStar Pneumocystis jirovecii PCR Kit and the AmpliSens Pneumocystis jirovecii (carinii)-FRT PCR Kit for Detecting P. jirovecii in Non-HIV Immunocompromised Patients.

Huh Hee Jae, Lim Kyoung Ree, Ki Chang Seok, Huh Kyungmin, Shim Hyang Jin, Song Dong Joon, Kim Yae Jean, Chung Doo Ryeon, Lee Nam Yong

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Real-time PCR is more sensitive than microscopic examination for detecting Pneumocystis jirovecii. We compared the performance of two assays for detecting P. jirovecii DNA: the RealStar Pneumocystis jirovecii PCR Kit 1.0 CE (Altona Diagnostics, Hamburg, Germany) and the AmpliSens Pneumocystis jirovecii (carinii)-FRT PCR kit (InterLabService Ltd., Moscow, Russia).

METHODS

We used 159 samples from the lower respiratory tract (112 bronchoalveolar lavage [BAL] fluid, 37 sputum, and 10 endotracheal aspirate [ETA] samples) of non-HIV immunocompromised patients. Nested PCR and sequencing were used to resolve discordant results. The performance of the two assays was evaluated according to clinical categories (clinical Pneumocystis pneumonia [PCP], possible PCP, or unlikely PCP) based on clinical and radiological observations.

RESULTS

The positive and negative percent agreement values were 100% (95% confidence interval [CI], 85.4-100%) and 96.6% (95% CI, 90.9-98.9%), respectively, and kappa was 0.92 (95% CI, 0.84-0.99). P. jirovecii DNA load was significantly higher in the clinical PCP group than in the other groups (P<0.05). When stratified by sample type, the positive rate for BAL fluids from the clinical PCP group was 100% using either assay, whereas the positive rate for sputum/ETA samples was only 20%.

CONCLUSIONS

The two assays showed similar diagnostic performance and detected low P. jirovecii burden in BAL fluids. Both assays may be useful as routine methods for detecting P. jirovecii DNA in a clinical laboratory setting, though their results should be interpreted considering sample type.

Для цитирования: Huh Hee Jae, Lim Kyoung Ree, Ki Chang Seok, Huh Kyungmin, Shim Hyang Jin, Song Dong Joon, Kim Yae Jean, Chung Doo Ryeon, Lee Nam Yong. Comparative Evaluation Between the RealStar Pneumocystis jirovecii PCR Kit and the AmpliSens Pneumocystis jirovecii (carinii)-FRT PCR Kit for Detecting P. jirovecii in Non-HIV Immunocompromised Patients.. Annals of laboratory medicine. 2019. DOI: 10.3343/alm.2019.39.2.176.