DOI 10.1016/j.cmpb.2020.105517

Медицинская библиотека · Computer methods and programs in biomedicine · 2020

Ultrasound diaphragmatic manual and semi-automated motion measurements: Application in simulated and in vivo data of critically ill subjects.

Loizou Christos P, Chrysostomou Constantinos, Minas Giorgos, Pattichis Constantinos S

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

Ultrasound diaphragmatic muscle motion characteristics may provide useful information about normal or abnormal diaphragmatic function and indicate diaphragmatic weakness, or paralysis. In the present work we propose and evaluate an integrated semi-automated analysis system for the quantitative analysis of ultrasonic motion from ultrasound diaphragmatic videos.

METHODS

The proposed system was evaluated in simulated videos and in 13 patients, four of whom patients were mechanically ventilated. The major steps of the methodology were as follows: video normalization, despeckle filtering, generation of an M-Mode image, snakes segmentation, and motion measurements.

RESULTS

The following manual (-/) vs semi-automated (/-), (median±IQR) measurements, which are routinely carried out by the experts, for assessing the severity of the disease, were computed. For the simulated videos the diaphragmatic excursion was 1.80±0.00 cm / 1.76±0.03 cm. For all the real ultrasound videos investigated in this study the following measurements were computed: (i) diaphragmatic excursion: 0.84±0.15 cm / 0.83±0.14 cm, (ii) inspiration time (Tinsp): 0.71±0.18 sec / 0.70±0.15 sec, (iii) total breathing time for one cycle (Ttot): 1.71±0.37 sec / 1.67±0.37 sec, (iv) diaphragmatic curve slope: 1.29±0.36 cm/sec / 1.27±0.36 cm/sec, and (v) relaxation rate (RR): 0.82±0.17 cm/sec / 0.82±0.18 cm/sec.

CONCLUSIONS

Manual and semi-automated measurements were very close with non-statistical significant differences and strong correlations between them. It is anticipated that the proposed system might be useful in the clinical practice in the assessment and follow up of patients with diaphragmatic weakness or paralysis and aid in the separation of normal and abnormal diaphragmatic motion. Further validation and additional experimentation in a larger sample of videos and different patient groups is required.

Для цитирования: Loizou Christos P, Chrysostomou Constantinos, Minas Giorgos, Pattichis Constantinos S. Ultrasound diaphragmatic manual and semi-automated motion measurements: Application in simulated and in vivo data of critically ill subjects.. Computer methods and programs in biomedicine. 2020. DOI: 10.1016/j.cmpb.2020.105517.

Предметные рубрики MeSH