MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ijg.0000000000000134

Медицинская библиотека · Journal of glaucoma · 2016

Initial Clinical Experience With the CyPass Micro-Stent: Safety and Surgical Outcomes of a Novel Supraciliary Microstent.

Hoeh Helmut, Vold Steven D, Ahmed Iqbal K, Anton Alfonso, Rau Magda, Singh Kuldev, Chang David F, Shingleton Bradford J, Ianchulev Tsontcho

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate safety and clinical outcomes of a novel supraciliary device, the CyPass Micro-Stent, for surgical treatment of open-angle glaucoma when implanted in conjunction with cataract surgery.

PATIENTS AND

METHODS

Subjects (n=142) with open-angle glaucoma and cataract underwent combined phacoemulsification, with intraocular lens insertion, and microstent implantation into the supraciliary space of study eyes (n=167). Two analysis cohorts were prespecified based upon medicated baseline intraocular pressure (IOP): ≥21 mm Hg (cohort 1, n=65) or <21 mm Hg (cohort 2, n=102). Glaucoma medications were discontinued or tapered at surgery, and restarted at investigator discretion. The main postoperative outcome measures were adverse events, IOP changes, and number of IOP-lowering medications.

RESULTS

Mean±SD follow-up was 294±121 days. No major intraoperative or postoperative complications occurred. Preoperative baseline mean IOP was 20.2±6.0 mm Hg and mean number of IOP-lowering medications was 2.0±1.1. Cohort 1 showed a 35% decrease in mean IOP and a 49% reduction in mean glaucoma medication usage; cohort 2 demonstrated a 75% reduction in mean medication usage while maintaining mean IOP<21 mm Hg. For all eyes, mean IOP at 12 months was 15.9±3.1 mm Hg (14% reduction from baseline). Early and late postoperative IOP elevation occurred in 1.2% and 1.8% of eyes, respectively. Two subjects developed mild transient hyphema, and none exhibited prolonged inflammation, persistent hypotony, or hypotony maculopathy.

CONCLUSIONS

CyPass Micro-Stent implantation, combined with cataract surgery, resulted in minimal complications and reduced IOP and IOP-lowering medication use at 12 months postoperatively.

Для цитирования: Hoeh Helmut, Vold Steven D, Ahmed Iqbal K, Anton Alfonso, Rau Magda, Singh Kuldev, Chang David F, Shingleton Bradford J, Ianchulev Tsontcho. Initial Clinical Experience With the CyPass Micro-Stent: Safety and Surgical Outcomes of a Novel Supraciliary Microstent.. Journal of glaucoma. 2016. DOI: 10.1097/ijg.0000000000000134.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.