DOI 10.1097/iae.0b013e3181aa8e6c
Short-term safety of 23-gauge single-step transconjunctival vitrectomy surgery.
Аннотация
PURPOSE: The purpose of this study was to report the safety profile of the Alcon 23-gauge single-step preloaded transconjunctival vitrectomy system.
METHODS
This was a retrospective case series examining 118 eyes of 114 patients that underwent 23-gauge vitrectomy between July and November 2007. Patients were observed postoperatively for a minimum of 6 weeks. Main outcome measures were postoperative intraocular pressure (IOP), intraoperative complications, and postoperative adverse events.
RESULTS
The mean postoperative Day 1 IOP was 14.3 +/- 6.6 mmHg (median, 13 mmHg; range, 2-38 mmHg) and Month 1 IOP was 15.8 +/- 5.9 mmHg (median, 15 mmHg; range, 6-52 mmHg). Two eyes (1.7%) developed choroidal effusions. Four eyes (3.4%) developed intraoperative horseshoe tears. No retinal detachments or endophthalmitis occurred. Four eyes (3.4%) had postoperative hypotony. Intraoperative wound leak after cannula removal requiring a sclerotomy suture occurred in 38% of patients: 35 eyes requiring closure of 1 cannula site (29.6%), 8 eyes (6.8%) requiring closure of 2 cannula sites, and 2 eyes (1.7%) requiring closure of all 3 cannula sites.
CONCLUSION
The Alcon 23-gauge single step preloaded transconjunctival vitrectomy system has few safety issues. Sclerotomy leaks after cannula removal are common, with 38% of eyes requiring suture closure of at least 1 cannula site.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.