New Validated Signal-averaging-based Electrocardiography Method to Determine His-ventricle Interval.
Аннотация
BACKGROUND/AIM: The signal-averaging (SA) technique is used to record high-resolution electrocardiograms (HRECGs) showing cardiac micropotentials. We aimed to develop a non-invasive signal-averaging-based portable bedside device to determine His-ventricle interval.
PATIENTS AND
METHODS
After amplifying the HRECG recordings, signal duration and voltage can be measured up to four decimal precision. To validate our system, comparison of the invasively and non-invasively determined HV intervals has been performed in 20 patients.
RESULTS
Our workgroup has developed a system capable of displaying and measuring cardiac micropotentials on storable ECG. Neither related paired-sample T-test (p=0.263) nor Wilcoxon's non-parametric signed ranks test (p=0.245) showed significant deviations of the HV intervals. Furthermore, related paired-sample T-test showed strong correlation (corr=0.910, p<0.001) between HV intervals determined by electrophysiology (EP) and non-invasive measurements.
CONCLUSION
Our research group managed to assemble and validate an easy to use device capable of determining HV intervals even under ambulatory conditions.