Safety and Efficacy of a Novel Approach to Pulmonary Vein Isolation Using Prolonged Apneic Oxygenation.

Gabriels James K, Ying Xiaohan, Purkayastha Subhanik, Braunstein Eric, Liu Christopher F, Markowitz Steven M, Mountantonakis Stavros, Thomas George, Goldner Bruce, Willner Jonathan, Goyal Rajat, Ip James E, Lerman Bruce B, Carter Jane, Bereanda Nicola, Fitzgerald Meghann M, Anca Diana, Patel Apoor, Cheung Jim W

Многоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Improved ablation catheter-tissue contact results in more effective ablation lesions. Respiratory motion causes catheter instability, which impacts durable pulmonary vein isolation (PVI).

OBJECTIVES

This study sought to evaluate the safety and efficacy of a novel ablation strategy involving prolonged periods of apneic oxygenation during PVI.

METHODS

We conducted a multicenter, prospective controlled study of 128 patients (mean age 63 ± 11 years; 37% women) with paroxysmal atrial fibrillation undergoing PVI. Patients underwent PVI under general anesthesia using serial 4-minute runs of apneic oxygenation (apnea group; n = 64) or using standard ventilation settings (control group; n = 64). Procedural data, arterial blood gas samples, catheter position coordinates, and ablation lesion characteristics were collected.

RESULTS

Baseline characteristics between the 2 groups were similar. Catheter stability was significantly improved in the apnea group, as reflected by a decreased mean catheter displacement (1.55 ± 0.97 mm vs 2.25 ± 1.13 mm; P < 0.001) and contact force SD (4.9 ± 1.1 g vs 5.2 ± 1.5 g; P = 0.046). The percentage of lesions with a mean catheter displacement >2 mm was significantly lower in the apnea group (22% vs 44%; P < 0.001). Compared with the control group, the total ablation time to achieve PVI was reduced in the apnea group (18.8 ± 6.9 minutes vs 23.4 ± 7.8 minutes; P = 0.001). There were similar rates of first-pass PVI, acute PV reconnections and dormant PV reconnections between the two groups.

CONCLUSIONS

A novel strategy of performing complete PVI during apneic oxygenation results in improved catheter stability and decreased ablation times without adverse events. (Radiofrequency Ablation of Atrial Fibrillation Under Apnea; NCT04170894).

Для цитирования: Gabriels James K, Ying Xiaohan, Purkayastha Subhanik, Braunstein Eric, Liu Christopher F, Markowitz Steven M, Mountantonakis Stavros, Thomas George, Goldner Bruce, Willner Jonathan, Goyal Rajat, Ip James E, Lerman Bruce B, Carter Jane, Bereanda Nicola, Fitzgerald Meghann M, Anca Diana, Patel Apoor, Cheung Jim W. Safety and Efficacy of a Novel Approach to Pulmonary Vein Isolation Using Prolonged Apneic Oxygenation.. JACC. Clinical electrophysiology. 2023. DOI: 10.1016/j.jacep.2022.10.030.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.