Local Impedance Drop Predicts Durable Conduction Block in Patients With Paroxysmal Atrial Fibrillation.

García-Bolao Ignacio, Ramos Pablo, Luik Armin, S Sulkin Matthew, R Gutbrod Sarah, Oesterlein Tobias, I Laughner Jacob, Richards Elizabeth, Meyer Christian, Yue Arthur, Ullah Waqas, Shepherd Ewen, Das Moloy

СтатьяКлиническое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This analysis was performed to evaluate the transition of local impedance (LI) drop during pulmonary vein isolation (PVI) to durable block and mature lesion formation based on 3-month mapping procedures.

BACKGROUND

A radiofrequency catheter measuring LI has been shown to be effective for performing PVI in patients with paroxysmal atrial fibrillation. Previous analysis has demonstrated LI drop to be predictive of pulmonary vein segment conduction block during an atrial fibrillation ablation procedure.

METHODS

Fifty-eight patients who had undergone LI-blinded de novo PVI returned for a 3-month mapping procedure. PVI ablation circles were divided into 16 anatomic segments for classification (durable block or gap), and the median LI drop within segments with an interlesion distance of ≤6 mm was compared. A total of 51 data sets met the criteria for segmental analysis of LI performance.

RESULTS

At the 3-month procedure, PV connection was confirmed in at least 1 PV segment in 35 of the included patients. LI drop outperformed generator impedance drop as a predictor of durable conduction block (area under the receiver-operating characteristic curve: 0.79 vs 0.68; P = 0.003). Optimal LI drops were identified by left atrial region (anterior/superior: 16.9 Ω [sensitivity: 69.1%; specificity: 85.0%; positive predictive value for durable conduction block: 97.7%]; posterior/inferior:14.2 Ω [sensitivity: 73.8%; specificity: 78.3%; positive predictive value: 96.9%]). Starting LI before radiofrequency (RF) application was significantly different among healthy, gap, and mature scar tissue and was also a contributing factor to achieving an optimal LI drop (85.2% of RF applications with a starting LI of ≥110 Ω achieved the optimal regional drop or greater).

CONCLUSIONS

LI drop is predictive of durable PV segment isolation. Preablation starting LI is associated with the magnitude of LI drop. These findings suggest that a regional approach to RF ablation guided by LI combined with careful interlesion distance control may be beneficial in patients with paroxysmal atrial fibrillation (Electrical Coupling Information From the Rhythmia HDx System and DirectSense Technology in Subjects With Paroxysmal Atrial Fibrillation [LOCALIZE]; NCT03232645).

Для цитирования: García-Bolao Ignacio, Ramos Pablo, Luik Armin, S Sulkin Matthew, R Gutbrod Sarah, Oesterlein Tobias, I Laughner Jacob, Richards Elizabeth, Meyer Christian, Yue Arthur, Ullah Waqas, Shepherd Ewen, Das Moloy. Local Impedance Drop Predicts Durable Conduction Block in Patients With Paroxysmal Atrial Fibrillation.. JACC. Clinical electrophysiology. 2022. DOI: 10.1016/j.jacep.2022.01.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.