The correlation between local impedance drop and catheter contact in clinical pulmonary vein isolation use.

Yasumoto Koji, Egami Yasuyuki, Kawanami Shodai, Sugae Hiroki, Ukita Kohei, Kawamura Akito, Nakamura Hitoshi, Matsuhiro Yutaka, Tsuda Masaki, Okamoto Naotaka, Matsunaga-Lee Yasuharu, Yano Masamichi, Nishino Masami, Tanouchi Jun

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Local impedance (LI) drop during radiofrequency (RF) application is monitored to assess the lesion formation. Recently, a novel ablation catheter has been introduced to clinical setting, which is capable of monitoring LI and catheter contact parameters including contact force (CF) and contact angle (CA). This study aimed to clarify the correlation between LI drop and catheter contact parameters.

METHODS AND

RESULTS

This prospective study included 15 paroxysmal atrial fibrillation (AF) patients who underwent initial pulmonary vein isolation (PVI). First-pass encircling point-by-point PV ablation was performed by using a 4.5-mm irrigated ablation catheter, with monitoring LI, CF, and CA. RF energy was applied for 30 s at each site with 30 W. Stable ablation points were analyzed to examine the correlation between LI drop and catheter contact parameters. Among 903 ablation points, 499 stable ablation points (55.2%) were analyzed. CA showed good correlation with LI drop (ρ = 0.418, p < .001). Maximum CF, minimum CF, average CF, and initial CF all showed weak correlation with LI drop (ρ = 0.201, p < .001; ρ = 0.224, p < .001; ρ = 0.258, p < .001; and ρ = 0.212, p < .001, respectively). Multivariate analysis demonstrated that CA was an independent factor of LI drop among the catheter contact parameters (β = 0.139, 95% CI = 0.111-0.167, p < .001). The LI drop in the blocked segments was significantly higher than that in the electrical conduction gap segments (27.3 ± 9.8 vs. 19.6 ± 6.4 Ω, p < .001)

CONCLUSION

In clinical PVI use, both CF and CA were correlated with LI drop. More parallel CA could induce higher LI drop, which may lead to effective lesion formation.

Для цитирования: Yasumoto Koji, Egami Yasuyuki, Kawanami Shodai, Sugae Hiroki, Ukita Kohei, Kawamura Akito, Nakamura Hitoshi, Matsuhiro Yutaka, Tsuda Masaki, Okamoto Naotaka, Matsunaga-Lee Yasuharu, Yano Masamichi, Nishino Masami, Tanouchi Jun. The correlation between local impedance drop and catheter contact in clinical pulmonary vein isolation use.. Pacing and clinical electrophysiology : PACE. 2022. DOI: 10.1111/pace.14500.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.