Clinical chemistry and laboratory medicine · 2022

Analytical evaluation and bioclinical validation of new aldosterone and renin immunoassays.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Coulon Caroline, Lotierzo Manuela, Fesler Pierre, Roubille Camille, Badiou Stéphanie, Dupuy Anne Marie, Cristol Jean Paul

Аннотация

OBJECTIVES: Aldosterone and renin determinations play an important role in the etiological diagnosis of secondary hypertension. The analytical performances of new aldosterone and renin immunoassays on the Lumipulse G600II® system (Fujierbio) were investigated and compared with those of the iSYS® system (IDS) on patients concerned by medical investigations in a context of suspected or proven Primary aldosteronism.

METHODS: By using the Lumipulse® G Aldosterone and Renin assays we performed imprecision study, linearity and method comparison (n=107). Accuracy of this new renin assay was tested using the International Standard (WHO IS 68/356). We also assessed the equivalence of the different samples types (n=29).

RESULTS: The imprecision evaluation showed all CVs <3% and <6% for Lumipulse® G Aldosterone and Renin assays respectively. The linearity was excellent over the clinical range and the comparison with the iSYS® assays (n=79) showed a strong correlation (R2=1) despite a slight tendency to underestimation (bias of -17.53 pg/mL or 48.56 pmol/L for aldosterone and -15.395 pg/mL for renin). Moreover, the contingency studies based on diagnostic criteria showed that Lumipulse® G results lead to the same clinical diagnosis that iSYS® results. A clear correlation was obtained between EDTA and heparin plasma as well as with the serum for all range of measures.

CONCLUSIONS: The Lumipulse® G Aldosterone and Renin assays present performances compatible with a routine use in medical laboratories. The precise quantification in the low range can be of interest in some clinical contexts especially standing/laying tests. However, the standardisation against the WHO International Standard Renin would be advisable.