MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.healun.2007.10.015

Медицинская библиотека · The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation · 2008

The Papworth experience with the Levitronix CentriMag ventricular assist device.

Shuhaiber Jeffrey H, Jenkins David, Berman Marius, Parameshwar Jayan, Dhital Kumud, Tsui Steven, Large Stephen R

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The Levitronix CentriMag ventricular assist device (VAD) is a centrifugal pump designed for short-term extracorporeal support in cardiogenic shock. The aim of this study is to report our clinical experience with the Levitronix CentriMag for uni- and biventricular support.

METHODS

Between July 2004 and December 2006, 27 patients were supported using the Levitronix CentriMag device. Nineteen were male. Mean age was 47.9 (range 19 to 72) years. Indications for support at implantation were cardiogenic shock that included: end-stage heart failure and too ill to undergo transplantation, with questionable neurologic status (9 subjects); right ventricular failure after left VAD (LVAD) implantation (5 subjects); post-cardiotomy status (7 subjects); and acute donor graft failure after heart transplantation (6 subjects).

RESULTS

Post-VAD 30-day survival was 30% (8 patients). Mean support time was 11 days for all patients (range 1 to 51 days). Mean support time for 14 Levitronix biventricular VADs was 11 (range 1 to 51) days. Mean support time for 7 Levitronix LVADs was 13.7 (range 1 to 30) days. The highest survival rates after Levitronix support were after donor graft failure (50%) and after cardiotomy (42%). Levitronix right VAD (RVAD) support after long-term LVAD insertion incurred 100% hospital mortality. Of those who survived, 8 patients were discharged home after VAD support and remain alive to date. Two patients were bridged to primary and another bridged to repeat heart transplantation. Five patients were weaned to recovery. Re-operation for bleeding occurred in 8 patients, clinical evidence of cerebral thromboembolism in 3, overwhelming sepsis in 1, and aortic thrombus formation in 1. Clot formation in the tubing was observed in 1 patient, necessitating emergent replacement at bedside, which was successful.

CONCLUSIONS

The Levitronix CentriMag system is a reliable and facile temporary circulatory support system as a bridge to decision in patients with refractory acute cardiogenic shock.

Для цитирования: Shuhaiber Jeffrey H, Jenkins David, Berman Marius, Parameshwar Jayan, Dhital Kumud, Tsui Steven, Large Stephen R. The Papworth experience with the Levitronix CentriMag ventricular assist device.. The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation. 2008. DOI: 10.1016/j.healun.2007.10.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.